YZ Systems, Inc. 3101 Pollok Drive Conroe, Texas USA 77303 P: 936.788.5593 F: 936.788.5720
Page 44
DP-2010 TP ver.10012004
S
ECTION
11: S
YSTEM
T
ROUBLESHOOTING
figure 35
voltage pulse
-
+
Ter. #5
Ter. #3
5-24 VDC
(20 mSec duration min.)
B. Voltage Pulse Input:
move
mode
switch 4 to the on position, refer to
figure 35.
Depress
the test switch and
hold
. A green LED should illuminate.
When the voltage pulse input is
received at the count input (ter. #3)
the red LED will turn on and off and
the green LED will illuminate again.
This will normally occur very quickly
and give the appearance that the red
LED blinks on when the pulse input is
received and removed, figure 36.
(CEVQT[
5[UVGO
#RRTQXGF
/WVWCN
figure 36
Z-65 Timer Mode
If the Z-65 controller is to be operated in the timer
mode, it acts a a simple recycling timer. Set up is
. If a sample is not taken
when expected in this mode the following should assist
in restoring the sampler to proper operation.