DP-2010RX Version 06042001 ATEX Rev.
YZ Systems, Inc. • 3101 Pollok Drive • Conroe, Texas • USA • 77303 • P: 936.788.5793 • F: 936.788.5720
15
9.3 Adjust the pump volume adjustment knob to
the value used in the calculations in step 9.1.
9.4 Turn mode switch 1 to on.
9.5 Turn mode switch 2 to on.
9.6 Turn mode switch 3 to off.
9.7 Turn mode switch 4 to off.
9.8 Press the test button once to initiate the timer
sequence.
NOTE: For the Z-65 to operate the DP-2010
System, a minimum pressure differential of 3" of
water must exist between the high pressure and
low pressure ports of the DPS-2.
Number of turns open on the pump volume knob
3
6
12
Sample pump displacement per stroke
.1cc
.2cc
.4cc
System
Mutual
Approved
Factory
9. Sampler set-up - time-based sampling with the Optional DPS-2
Содержание DP-2010RX
Страница 1: ...DynaPak GasSampler S Y S T E M S U P P O R T M A N U A L DP 2010RX...
Страница 2: ...DynaPak2010RX SystemSupport Manual Version 06042001 ATEX Rev...
Страница 25: ...26 Diagram 1 DP 2000 pump assembled...
Страница 27: ...28 Diagram 3 YZ filter regulator assembled...
Страница 35: ...36 Notes...