1.4.2 Probe Connection
When you use YUT-2600, we recommend using our stock probe connecting cable with the
appropriate probes. Various types of our standard stock probes can be used directly.
The probe should be plugged into the socket upside the instrument. When it is the mode of single
probe, the probe cable should be connected to T/R socket; when it is the mode of dual probe (one for
transmitting, the other for receiving), you should plug the transmitting probe into the T/R socket and the
receiving probe into the R socket. If these factors are not paid attention to, unfavorable results might be
loss or disordered echo waveform.
1.4.3 Switching On and Off the Instrument
Press to directly start up YUT-2600 and activate the standby mode. After the detection and
data storage, press to shut down the instrument.
1.4.4 Restore the Initial Parameters of the Instrument
If the instrument crashes or some functions are in disorder or the parameters of the instrument need
resetting, press twice before press (up) key or (down) key to select the submenu
resetting parameters. Then press (left) key or (right) key to adjust the parameters. When the
screen displays resetting parameters confirmation, press Ok to restore the parameters.
Note: after the instrument is initialized, all the parameters are reset to the factory defaults.
1.5 Main Menu and Functions
1.5.1 Functional Keys
Automatic gain
Increase or decrease the highest wave inside the gate to 80% of the
screen
Freezing
Waveform freezing
Wave Peak Memory
Automatically store the highest wave
Full screen
Full screen display
On/off
Switch on and off
Menu control keys
Up, down, left, right,
Ok
Up/down keys are used to switch among options, left/right keys are
used to adjust values.