![YUSHI YUT-2800 Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/yushi/yut-2800/yut-2800_manual_3745659042.webp)
YUSHI INSTRUMENTS
39
continuous transfer.
Connect one end of the communication cable with the communication interface. The red pot of the
interface should be aligned with the red pot on the instrument. The other end of the communication cable
should be connected with the COM1 (or COM2) interface on the PC. The instrument will transmit the
saved graphs and data to the PC. The client software installed on PC can edit, save or print these graphs
and data.
Operating steps:
1. Connect the instrument and PC with the communication cable.
2. Press to enter the save menu. Use the up and down keys to select the communication
submenu. Use right and left keys to turn it on. Then the instrument is ready to connect the PC.
3. Open the communication software DDIclient.exe, and tap in the file scopes for download. Then
press the download button.
Note: 1. Pulling out the communication cable or shutting down the software during communicating will
fail the communication and lead to the breakdown of the instrument.
2. Please restart the instrument if the communication fails.
The software is programmed with C++. The language is Chinese. The picture below shows the
running of the software.
s