![YUSHI YUT-2800 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/yushi/yut-2800/yut-2800_manual_3745659025.webp)
YUSHI INSTRUMENTS
22
3.8.2 Freezing
The function can freeze and lock the display on the current screen. In the freezing mode, you can
move the gate to measure the echoed signal frozen on the screen.
If you want to freeze the current display, press . If you want to return the normal mode,
press a gain to unfreeze.
Note: in the freezing mode, all the adjustable parameters (gain, sound velocity) can not be adjusted.
3.8.3Wave Peak Memory
Wave peak memory is a function that the instrument compares the “new” with the “old” highest
dynamic echoes automatically and preserves the highest still echo.
First use the gate to lock the echo needed, press to activate the wave peak memory status.
When you move the probe, if there emerges a dynamic echo which is higher than the previous static
wave displayed, the instrument can catch the new echo as the current highest echo displayed. Press
again to exit the wave peak memory.
3.8.4 Full Screen
Press to enlarge the waveform display area to full screen so that you can observe the echo
and the measurement values conveniently. At this moment, the main menu display is canceled. If you
want to return to the normal display mode, press again.
Note: in full screen mode, except the functions of automatic gain, wave peak memory, and freezing
are available, other functions will be locked and unavailable.
Chapter 4 Application of Testing
4.1 Calibration Method and Steps of the Instrument
Before you use YUT-2600 to test the work piece, you should first calibrate the instrument:
according to the material, size of the work piece to be tested as well as the relevant standards, choose the