background image

Blanking in the dual-echo mode:

1. Initial-blanking: red blanking line indicated on the screen, starting at
zero, so named initial blanking. Waveform with the scope of red strip is
invalid, for omitted the clutter between the starting point and the first
echo.
2. E-blanking (echo-blanking): blue blanking line indicated on the
screen, only appearance when successful measuring. Starting at the
first echo measurement point, so named echo-blanking. Waveform
within the scope of blue stripe is invalid, for omitted the clutter between
the first echo and secondary echo.

3.8 Interface wave-Echo Mode

Interface wave mode makes use of interface wave and echo to
measure the thickness of the work piece. This method has wide-range
of measuring, the upper limit can reach to 27mm.

3.9 Multiple-wave check method(High Precision Mode)

PM-5 series provides Multiple-wave check method, the measurement
range is 0.2mm-10mm. According to 3(Min. times) to 9Max.times) times
echos every time when measuring, the more measuring times, the
higher precision will get. Furthermore the reading values are all through
checkout, to make sure the results are accurate.We recommend to use
this mode when measuring super thin and high precision work piece.

3.9.1 Illustration of Multiple-wave check mode

3.17 Multiple-Wave Check Mode

1

Identification of multiple-wave mode 2

red arrow indicates the

first echo that participates in verifying 3

blue arrow indicates the last

echo that participates in verifying 4

M indicates the numbers the

participates in verifying
This picture shows A-scan snapshot wave of measuring the wall
thickness of fire extinguisher under Multiple-wave check mode. The wall
of the fire extinguisher is very thin and has paint coating as well, which
make it difficult to measure by using the normal thickness gauge.
Through using Multiple-wave check mode of PM-5DL, we can see from
the picture that there are 9 verifying waves to make sure the measuring
result accurate and stable.

3.10 AUTO Mode

The unique Auto mode of PM-5 series measurement range is from
0.2mm-27mm. On the basis of thickness, it selects the measuring
model automatically. When H≤10mm, it adopts ME-E measuring mode;
when 0.2mm<H<=15mm, it adopts I-E measuring mode; and when
H>2mm, it adopts P-E mode. Under this mode, the user only needs to
adjust the Gain. It is easy to operate, we recommend the common users
to adopt this mode. H is the thickness of the material to be measured

4.Date Storage Function

PM-5DL

PM-5DL adopt the storage mode of micro-grid format, it can save one
hundred thousand thickness values and one thousand A/B-scan
wave-forms, waveform and thickness values can be mixed stored in the
same file. The measurement data files can be transferred from the
instrument to a PC via USB communication to generate EXCEL or TXT
format files. Using our powerful Data View software to statistical and
analyze measurement, report via connect printer.

21

22

Содержание PM-5 Series

Страница 1: ... 3 5 Example Explanation of A SCAN 17 3 6 Operation of B Scan Interface 19 3 6 1 B Scan Display 19 3 6 2 Introduction of B Scan 19 3 7 Dual Echo Thru Thin Coating Mode 20 3 7 1 A scan Interface In Dual Echo Mode 20 3 8 Interface wave Echo Mode 21 3 9 Multiple wave check method High Precision Mode 21 3 9 1 Illustration of Multiple wave check mode 21 3 10 AUTO Mode 22 4 Date Storage Function 22 4 1T...

Страница 2: ...nhance the accuracy makes the resolution reach to 0 001mm and the lower limit as little as 0 2mm In addition A B scan waveform display is available by showing the echoes which enables the more extensive measure range more precise result and very little measurement error and finally makes the measuring processes more controllable This series of Instrument can be widely used to measure the thickness...

Страница 3: ...rm Power Requirements 2 AA size batteries Operating Time Greater than 35 hours Instrument Shut off Selectable ALWAYS ON or AUTO OFF after 5 10 20 minutes of inactivity Operating Temperature 10 C to 50 C 10 F to 120 F 20 C for special requirements Size 153mm 76mm 37mm H W D Weight 280g including batteries Warranty One year 1 5 Main Functions 1 Unique Multiple wave check method make use of multiple ...

Страница 4: ... mm and X X mm X XXXX inch X XXX inch and X XX inch 12 Optional waveform style outline mode or fill mode 13 Optional rectification mode RF RF full wave half half 14 Support adjustment of velocity and single point calibration 15 Multi languages Available Chinese English German French and Japanese 16 Approx battery life 35 hours 17 Great capacity data storage function Stores 100 000 thickness values...

Страница 5: ...nown velocity of the material e g The velocity of the steel is 5900 s then input the velocity manual 3 1 1 One Point Calibration When the sound velocity of the material to be measured is unknown sound velocity measurement must be made before measuring the thickness Prepare a test block with the same composition as the object to be measured and its surface must be suitable for measurement Measure t...

Страница 6: ...asuring the more measuring times the higher precision will get Furthermore the reading values are all through checkout to make sure the results are accurate AUTO Automatic mode measurement range is from 0 2mm 27mm On the basis of thickness it selects the measuring model automatically When H 10mm it adopts ME E measuring mode when 0 2mm H 5mm it adopts E E measuring mode and when H 2mm it adopts P ...

Страница 7: ...mode A scan interface B scan interface And there are three display modes of normal interface Thickness value mode Difference rate of reduction measurement mode MAX MIN measurement mode Select in the VIEW MODE of CONFIGURATION When the probe and the object are not completely coupled the letters in the various interfaces are in GREEN when properly coupled they are displayed in WHITE color and when t...

Страница 8: ... value 3 respectively are battery power probe type degree of gain Pulse Echo measuring mode and measuring unit 4 A scan interface 5 reset 6 the current thickness value battery power 7 the maximum value 8 file number 3 4 A Scan Interface Mode In this mode users could view the current thickness value and the A scan waveform at the same time The right side of the interface is the specification adjust...

Страница 9: ...if the delay is set incorrectly and the echo waveform is beyond the display area however the testing value can also be read correctly BLANK Hide the unnecessary and useless clutter in front of the main waves The red line on the bottom of the screen shows the blank confines The adjusting blank confines are the present range confines In the real testing wrong measurement due to the material may exis...

Страница 10: ...ure the thickness of the testing object is about 10mm but as there are obvious defects the defect echoes are showed on the display and the gate locks the defect echoes which have broken the gate thus the testing result shown is the thickness of the defect area The right measurement can be realized by adjusting the gate setting above the defect echoes 3 13 Real Case 2 3 If there are some surface fa...

Страница 11: ... Dual Echo Thru Thin Coating Mode For the thin protect coating and paint the thickness measurement of the underlying metal will lead significant error when using conventional thickness gauge PM 5DL can accurate measure the actual thickness of the underlying metal with dual echo measurement principle and without having to remove the coatings or destroy the surface process This function is achieved ...

Страница 12: ...he first echo that participates in verifying 3 blue arrow indicates the last echo that participates in verifying 4 M indicates the numbers the participates in verifying This picture shows A scan snapshot wave of measuring the wall thickness of fire extinguisher under Multiple wave check mode The wall of the fire extinguisher is very thin and has paint coating as well which make it difficult to mea...

Страница 13: ...ections In this condition the preset velocity should be the average value among the testing range which can be acquired through testing a block with the same velocity as the object average velocity value 2 ULTRA THIN MATERIAL When the thickness of the testing object is below the minimum value of the probe limit the result may be incorrect and the thickness can be acquired by contracting the blocks...

Страница 14: ...me please remove batteries to prevent any leakage or corrosion 6 2 Considerations 1 Please be cautious of the zero block s getting rust as couplant will be spread on the surface of it when calibrating the gauge After using clean the zero block Avoid dripping sweat on the gauge in high temperature Some grease spreading on the surface of zero block is useful to avoid rusting if the gauge will be spa...

Страница 15: ...0 180 4600 Lead 0 085 2200 Magnesium 0 230 5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0 067 1700 Platinum 0 130 3300 Plexiglass 0 110 1700 Polyethylene 0 070 1900 Polystyrene 0 0930 2400 Polyurethane 0 0700 1900 Quartz 0 230 5800 Rubber Butyl 0 070 1800 Silver 0 140 3600 Steel Mild 0 233 5900 Steel Stainless ...

Отзывы: