EN
30
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The cus
-
tomer or the user of the device should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted
RF
IEC 61000-
4-6
Radiated RF
IEC 61000-
4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter. Recommended
separation distance
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 80 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
Distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnet
-
ic site survey*, should be less than the compliance
level in each frequency range.** Interference may
occur in the vicinity of equipment marked with the
following symbol:
Notes:
1. At 80 MHz and 800 MHz, the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
*. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele
-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the IH25/1 is used exceeds the applicable RF compliance level above, the device
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
**. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Содержание ULAIZER HOME CN-02 MY
Страница 1: ...ІНСТРУКЦІЯ КОРИСТУВАЧА DE UA PL EN RO IT RU ES КОМПРЕСОРНИЙ НЕБУЛАЙЗЕР ULAIZER HOME CN 02 MY ...
Страница 10: ...USER S MANUAL COMPRESSOR NEBULIZER ULAIZER HOME CN 02 MY ...
Страница 18: ...MANUALE UTENTE NEBULIZZATORE A COMPRESSORE ULAIZER HOME CN 02 MY ...
Страница 26: ...MANUAL DE USUARIO NEBULIZADOR DE COMPRESOR ULAIZER HOME CN 02 MY ...
Страница 34: ...BENUTZERHANDBUCH KOMPRESSOR VERNEBLER ULAIZER HOME CN 02 MY ...
Страница 42: ...INSTRUKCJA OBSŁUGI NEBULIZATOR KOMPRESOROWY ULAIZER HOME CN 02 MY ...
Страница 50: ...MANUAL DE UTILIZARE NEBULIZATOR CU COMPRESOR ULAIZER HOME CN 02 MY ...
Страница 58: ...ИНСТРУКЦИЯ ПОЛЬЗОВАТЕЛЯ КОМПРЕССОРНЫЙ НЕБУЛАЙЗЕР ULAIZER HOME CN 02 MY ...