Calibration
Section 5
YSI, Incorporated
Model 85
13
10.00
µ
S
24.8
o
C
CAL
9.
Use the
UP ARROW
or
DOWN ARROW
button to adjust the reading on the display until it
matches the value of the calibration solution you are using.
10.
Once the display reads the exact value of the calibration solution being used (the instrument will
make the appropriate compensation for temperature variation from 25°C), press the
ENTER
button once. The word “
SAVE”
will flash across the display for a second indicating that the
calibration has been accepted.
The YSI Model 85 is designed to retain its last conductivity calibration permanently. Therefore,
there is no need to calibrate the instrument after battery changes or power down.
Содержание 85
Страница 2: ...YSI Model 85 Handheld Oxygen Conductivity Salinity mg L C and Temperature System Operations Manual ...
Страница 6: ...Introduction Section 1 YSI Incorporated Model 85 2 ...
Страница 10: ...Preparing the Meter Section 2 YSI Incorporated Model 85 6 ...
Страница 12: ...Preparing the Probe Section 3 YSI Incorporated Model 85 8 ...
Страница 14: ...Overview of Operation Section 4 YSI Incorporated Model 85 10 ...
Страница 18: ...Calibration Section 5 YSI Incorporated Model 85 14 ...
Страница 23: ...Making Measurements Section 7 YSI Incorporated Model 85 19 ...
Страница 24: ...Making Measurements Section 7 YSI Incorporated Model 85 20 ...
Страница 28: ...Maintenance Section 9 YSI Incorporated Model 85 24 ...
Страница 36: ...Discussion of Measurement Errors Section 11 YSI Incorporated Model 85 32 ...
Страница 43: ...Warranty and Repair Section 13 YSI Incorporated Model 85 39 ...
Страница 45: ...Accessories and Replacements Section 14 YSI Incorporated Model 85 41 ...
Страница 47: ...Specifications Appendix A YSI Incorporated Model 85 43 ...
Страница 53: ...Required Notice Appendix C YSI Incorporated Model 85 49 ...
Страница 55: ...Conversion Chart Appendix D YSI Incorporated Model 85 51 ...
Страница 59: ...Calibration Values Table Appendix F YSI Incorporated Model 85 55 ...