Advanced Conductivity Setup
Section 6
YSI, Incorporated
Model 85
16
6.2 CHANGING THE REFERENCE TEMPERATURE
Follow these steps to modify the reference temperature of the Model 85.
1.
Turn the instrument on and wait for it to complete its self-test procedure.
2.
Use the
MODE
button to advance the instrument to display conductivity.
3.
Press and release both the
DOWN ARROW
and the
MODE
buttons at the same time.
The
CAL
symbol will appear at the bottom left of the display. The large portion of the display will
show
1.91 %
(or a value set previously using Advanced Setup).
4.
Press and release the
MODE
button; the large portion of the display will show
25.0C
(or a
value set previously using Advanced Setup).
5.
Use the
UP ARROW
or
DOWN ARROW
button to change the value to the desired new
reference temperature (any value between 15
o
C and 25
o
C is acceptable).
6.
Press the
ENTER
button. The word “
SAVE
” will flash across the display for a second to
indicate that your change has been accepted.
7.
The instrument will automatically return to normal operation mode.
6.3 CHANGING FROM AUTORANGING TO MANUAL RANGING
If your application is easier to perform using a manual range that you select, the YSI Model 85
allows you to turn off the default autoranging feature. While you are making conductivity or
temperature compensated conductivity measurements, simply press and release the
UP ARROW
button. Each additional press of the
UP ARROW
button will cycle the Model 85 to a different
manual range until you return again to autoranging. Five pushes of the
UP ARROW
button will
cycle the Model 85 through the four manual ranges and return the instrument to autoranging.
NOTE:
You may see an error message in some manual ranges if the manual range selected is not
adequate for the sample you are measuring. If this happens, simply press and release the
UP
ARROW
button again until a range is selected which is suitable for your sample. If you get lost and
don’t know if you’re in a manual range or autoranging, simply turn the instrument off and back on.
Also note that the conductivity units will flash while you are in manual range. The instrument will
always default to autoranging when first turned on.
The four ranges of the YSI Model 85 are:
Range 1
Range 2
Range 3
Range 4
0 to 499.9
µ
S/cm
0 to 4999
µ
S/cm
0 to 49.99 mS/cm
0 to 200.0 mS/cm
Содержание 85
Страница 2: ...YSI Model 85 Handheld Oxygen Conductivity Salinity mg L C and Temperature System Operations Manual ...
Страница 6: ...Introduction Section 1 YSI Incorporated Model 85 2 ...
Страница 10: ...Preparing the Meter Section 2 YSI Incorporated Model 85 6 ...
Страница 12: ...Preparing the Probe Section 3 YSI Incorporated Model 85 8 ...
Страница 14: ...Overview of Operation Section 4 YSI Incorporated Model 85 10 ...
Страница 18: ...Calibration Section 5 YSI Incorporated Model 85 14 ...
Страница 23: ...Making Measurements Section 7 YSI Incorporated Model 85 19 ...
Страница 24: ...Making Measurements Section 7 YSI Incorporated Model 85 20 ...
Страница 28: ...Maintenance Section 9 YSI Incorporated Model 85 24 ...
Страница 36: ...Discussion of Measurement Errors Section 11 YSI Incorporated Model 85 32 ...
Страница 43: ...Warranty and Repair Section 13 YSI Incorporated Model 85 39 ...
Страница 45: ...Accessories and Replacements Section 14 YSI Incorporated Model 85 41 ...
Страница 47: ...Specifications Appendix A YSI Incorporated Model 85 43 ...
Страница 53: ...Required Notice Appendix C YSI Incorporated Model 85 49 ...
Страница 55: ...Conversion Chart Appendix D YSI Incorporated Model 85 51 ...
Страница 59: ...Calibration Values Table Appendix F YSI Incorporated Model 85 55 ...