Surges
IEC 61000-4-5
±0.5kV, ±1kV
lines to lines;
±0.5kV, ±1kV, ±2kV
lines to earth
±0.5kV, ±1kV
lines to lines;
Mains power quality should be that
of a HOME HEALTHCARE
ENVIRONMENT or a Professional
healthcare
facility environment..
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V
0.15MHz – 80MHz,
6V in ISM bands
between 0.15MHz and
80MHz (Professional
healthcare facility
environment), 6V in ISM
and amateur radio
bands between
0.15MHz and 80MHz
(Home healthcare
environment)
80% AM at 1kHz
Applicable
Note: The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795
MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur
radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz,
7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4
MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Rated power frequency
magnetic fields
IEC 61000-4-8
30A/m
50Hz or 60Hz
30A/m
50Hz
Power frequency magnetic fields
should be at levels characteristic
of a HOME HEALTHCARE
ENVIRONMENT or a
Professional healthcare
facility environment.
Voltage dips
IEC 61000-4-11
0% U
T
, 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°;
0% U
T
, 1 cycle
and
70% U
T
, 25/30 cycle
Single phase: at 0°
Applicable
Mains power quality should be
that of a HOME HEALTHCARE
ENVIRONMENT or a
Professional healthcare
facility environment...
Voltage interruptions
IEC 61000-4-11
0% U
T
, 250/300 cycle
Applicable