in nearby electronic equipment.
Conducted RF emissions
CISPR 11
Group 1
Class B
The
monitor
is suitable for use in all establishments,
including domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes except for near
active HF surgical equipment and the RF shielded room
for magnetic resonance imaging.
Radiated RF emissions
CISPR 11
Group 1
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations and flicker
emissions
IEC 61000-3-3
Applicable
Guidance and manufacture’s declaration – electromagnetic immunity
The upper arm blood pressure monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the upper arm blood pressure monitor should assure that it is used in such an
environment.
Immunity test
IEC 60601-1-2 test
level
Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge
IEC 61000-4-2
±8kV contact;
±2kV, ±4kV, ±8kV, ±15
kV air
±8kV contact;
±2kV, ±4kV,
±8kV, ±15kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Radiated RF EM
fields
IEC 61000-4-3
3V/m
(Professional healthcare
facility environment);
10V/m
(Home healthcare
environment),
80MHz – 2.7GHz 80%
AM at 1kHz
10V/m
(Home healthcare
environment)
80MHz – 2.7GHz 80% AM
at 1kHz
Electrical fast
transients/bursts
IEC 61000-4-4
±2kV AC power supply
lines;
±1kV DC power/Signal
lines.
100 kHz repetition
frequency
±2kV AC power supply
lines;
Mains power quality should be that
of a HOME HEALTHCARE
ENVIRONMENT or a Professional
healthcare
facility environment.