![YOKOHAMA ADMAG TI AXFA11G Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/yokohama/admag-ti-axfa11g/admag-ti-axfa11g_user-manual_911764063.webp)
Warning
The device works normally and measurement is also normal but warning occurs.
NE107
Status
Error Message
Error Description
Countermeasure
Display
S
080:AO 1 saturate
Saturation of Analog output 1 was detected.
Check process value and current output 1 parameter setting.
S
081:AO 2 saturate
Saturation of Analog output 2 was detected.
Check process value and current output 2 parameter setting.
S
082:Pls 1 saturate
Saturation of Pulse output 1 was detected.
Check process value and pulse output 1 parameter setting.
S
083:Pls 2 saturate
Saturation of Pulse output 2 was detected.
Check process value and pulse output 2 parameter setting.
S
084:AI saturate
Saturation of Analog input was detected.
Check current input value and current input parameter setting.
C
085:Cable miscon
Misconnection of cable was detected.
Check the signal cable and excitation cable connection.
C
086:Coil insulation
Insulation deterioration of coil was detected.
Contact Yokogawa service center.
M
087:Adhesion lv 3
The resistance value of electrode exceeded Level 3.
(Adhesion detection of insulation to electrode)
Recommend cleaning electrode.
N
088:LC warn
Decrease of conductivity was detected.
Check fluid conductivity.
M
089:Insu detect
Insulation deterioration of electrode was detected.
Contact Yokogawa service center.
C
092:AZ warn
Result of Autozero adjustment |Mag Flow Zero| exceeded
10 cm/s.
Check fluid is stopped when executing auto-zero adjustment.
C
093:Verif warn
Interruption of verification function was detected.
Execute Verification again.
C
094:Fact noise warn
Fluctuation of flow became larger.
Check the fluid.
C
095:Simulate active
Test mode was executed for any of Flow velocity,
Volumetric flow rate, Mass flow rate, Calorie, Current
output, Pulse output, Status input, Status output.
Release simulation or test mode.
S
096:AO 1 fix
It was detected that fixied value is set to Current output 1.
Check current output 1 is test mode or not.
S
097:AO 2 fix
It was detected that fixied value is set to Current output 2.
Check current output 2 is test mode or not.
S
098:Pls 1 fix
It was detected that fixied value is set to Pulse output 1.
Check pulse output 1 is test mode or not.
S
099:Pls 2 fix
It was detected that fixied value is set to Pulse output 2.
Check pulse output 2 is test mode or not.
S
100:AI fix
It was detected that fixied value is set to Current input.
Check current input is test mode or not.
C
101:Param restore run Restore function of parameter is running.
―
N
102:Disp over
Number of digits available for display exceeded the limit.
Check Display format parameter.
N
103:SD size warn
Free space of microSD card decreased to less than 10%.
microSD card may run out of memory space.
M
104:Bkup incmplt
Parameter backup failed.
Retry parameter backup.
S
105:SD mismatch
Mismatch of microSD card was detected.
Change microSD card.
M
106:SD removal ERR
Removal of microSD card failed.
Remove microSD in appropriate rocedure.
N
131:Trans mismatch
Mismatch of sensor and transmitter was detected.
Contact Yokogawa service center.
Information
The device works normally and measurement is also normal. Just reference information.
NE107
Status
Error Message
Error Description
Countermeasure
Display
N
120:Watchdog
Error of Watchdog timer was detected.
Contact Yokogawa service center.
N
121:Power off
Power-off was detected.
―
N
122:Inst power FAIL
Instantaneous power failure was detected.
―
N
123:Param bkup run
Parameter backup is running.
―
N
124:Data log run
Data log is running.
―
N
130:DevID not enter
No data entry of Device ID was detected.
Contact Yokogawa service center.
N
132:Modbus cfgupdate
The Modbus configuration has been changed, but the
device is not operating according to its configuration.
Restart Modbus communication.
N
133:G/A mismatch
As a result of self-test, inconsistency of internal circuit was
detected.
Contact Yokogawa service center.
N
246:Simulate SW OFF
Writing to parameters is locked.
Change the RB Write Lock (RB. WRITE_LOCK) to Not Locked or
turn off the hardware write lock switch, depending on RB Feature
Selection (RB.FEATURE_SEL).
N
248:Simulate SW ON
Software or hardware simulation switch is ON.
Delete the value of MTB Sim Enable Message (MTB.SIM_ENABLE_
MSG) or turn off the hardware simulation switch.
<7. Errors and Countermeasures (Display unit)>
62