<5. Basic Operating Procedures>
47
IM 01E25A01-01EN
5.4.1 Setting example of Select type
Data: Flow rate unit
The following is the procedure of changing the flow rate
unit as Select type parameter.
The flow rate unit needs to be specified with “Physical
unit” and “Time unit” individually. When the flow rate unit
needs to be set “l/min”, select “l (litter)” at the Physical unit
and “/min” at the Time unit.
Display Menu Path:
Device setup ► Detailed setup ► Pro var ► Volume ► Unit
Device setup ► Detailed setup ► Pro var ► Volume ► Time Unit
F0526.ai
Volume
Damp pls/ttl
Unit
Time unit
▲
▼
00 : 00
SET
SFT
INC
Specify the Physical unit for volumetric flow.
Move cursor with [INC] and [DEC] according
to the menu path above, and select the “unit”
then push [SET].
F0527.ai
Unit
m
3
l(liter)
cm
3
▲
▼
00 : 00
SET
SFT
INC
Move cursor with [INC] and [DEC], and
select the “l(liter)” then push [SET]. As the
selected unit is blinking, push [SET] to
determin. The screen returns to the setting
page.
F0528.ai
Volume
Unit
Time unit
Span
▼
00 : 00
SET
SFT
INC
▲
Specify the Time unit for volumetric flow.
Move cursor with [INC] and [DEC] according
to the menu path above, and select the “Time
unit” then push [SET].
F0529.ai
Time unit
/s
/min
/h
▼
00 : 00
SET
SFT
INC
▲
Move cursor with [INC] and [DEC], and
select the “/min” then push [SET]. As the
selected unit is blinking, push [SET] to
determin. The screen returns to the setting
page after the setting.
F0530.ai
0.00000m/s
0.00000
l/min
00 : 00
PRV
VEL
FLB
SET
SFT
INC
After completing the parameter setting, push
[ESC] then the screen returns to the Display
Mode.
NOTE
Be sure to set the Flow rate unit in the beginning when
the Flow rate unit and Flow span value are changed at
the same time.
When the unit is changed, the value of flow rate span is
converted to related values automatically according to
the unit change.
5.4.2 Setting example of Numeric type
Data: Flow rate span
The following is the procedure of changing the Flow rate
span as Numeric type parameter.
Display Menu Path:
Device setup ► Detailed setup ► Pro var ► Volume ► Span
NOTE
Be sure to set the Flow rate unit in the beginning when
the Flow rate unit and Flow span value are changed at
the same time.
When the unit is changed, the value of flow rate span is
converted to related values automatically according to
the unit change.
F0531.ai
Volume
Time unit
/h
Span +7200.00
Setting download
00 : 00
SET
SFT
INC
▲
▼
Specify the Flow rate span unit.
Move cursor with [INC] and [DEC] according
to the menu path above, and select the
“Span” then push [SET].
F0532.ai
Span
+7200.00
+2200.00
+0.00001 +999999
00 : 00
SET
SFT
INC
The switch's functionality of setting the Flow
rate span is as below:
Plus/minus and mumeric change: [INC]
Movement on digits: [SFT]
Determination of parameter: [SET]
L
: Minimum value
U
: Maximum value
Push [SET] to decide while the value of Flow
rate span is blinking. The screen returns to
the setting page after the setting.
5.4.3 Setting Example of Alphanumeric type
Data: Tag No.
The following is the procedure of changing the Tag No. as
Alphanumeric type parameter.
Display Menu Path:
Device setup ► Detailed setup ► Device info ► Order info ► Tag No.
F0533.ai
Device info
Date/Time
Order info
Ver/Num info
00 : 00
SET
SFT
INC
▲
▼
►
►
►
Specify the Tag No..
Move cursor with [INC] and [DEC] according
to the menu path above, and select the
“Order info” then push [SET].
F0534.ai
Tag No
FT-1234
FT-1234
00 : 00
SET
SFT
INC
For Tag No., up to 8 characters can be
entered from display panel.
The switch's functionality of setting the Flow
rate span is as below:
Plus/minus and mumeric change: [INC]
Movement on digits: [SFT]
Determination of parameter: [SET]
Available characters: ASCII characters
Push [SET] to decide while the value of Tag
No. is blinking. The screen returns to the
setting page after the setting.
Basic Operating Procedures