<4. Wiring>
27
IM 01E25A01-01EN
Remote Transmitter
AXFA11G
Optional
Code
Description
EW
5 pcs. of waterproof gland with JIS G3/4 female
EG
5 pcs. of waterproof gland
EU
5 pcs. of waterproof gland with union joint
EP
5 pcs. of plastic gland
IMPORTANT
Apply a blanking plug to the unused electrical
connection. Seal the electrical connections properly
comply to usage state.
4.3.1 When waterproof property is neces
-
sary (Wiring using waterproof glands)
IMPORTANT
To prevent water or condensation from entering the
transmitter housing, waterproof glands are recommended.
Do not over-tighten the glands or damage to the cables
may result. Tightness of the gland can be checked by
confirming that the cable is held firmly in place.
For working on the conduit or the flexible tubes (G1/2),
remove the waterproof gland and attach them directly to
the electrical connection.
(1) Waterproof Gland
Integral Flowmeter
Optional code: EG2, EG3
Cable
Waterproof gland
Gasket
Washer
F0405.ai
Remote Sensor
Optional code: EG
Cable
Waterproof gland
Gasket
Washer
F0406.ai
Remote Transmitter (AXW4A)
Optional code: EG4, EG5
Cable
Waterproof gland
Gasket
Washer
F0407.ai
Remote Transmitter (AXG1A/AXFA11)
Optional code: EG5 (AXG1A), EG (AXFA11)
F0408.ai
Cable
Waterproof gland
Gasket
Washer
W
iring
4