3. Installation
WARNING
Installation of the magnetic flowmeter must be
performed by expert engineer or skilled personnel.
No operator shall be permitted to perform procedures
relating to installation.
Installation Location Precautions
Select the installation location with consideration to the
following items to ensure long-term stable operation of
the instrument.
Ambient Temperature:
Avoid installing the instrument in locations with constantly
fluctuating temperatures. If the location is subject to
radiant heat from the plant, provide heat insulation or
improve ventilation.
Atmospheric Condition:
Avoid installing the instrument in a corrosive atmosphere.
In situations where this is unavoidable, consider ways to
improve ventilation and to prevent rainwater from entering
and being retained in the conduit pipes.
Vacuum:
In the case of PTFE lining, avoid the negative pressure
inside the measuring pipe.
Vibrations or Shocks:
Avoid installing the instrument in a place subject to
shocks or vibrations.
3.1 Piping Design Precautions
IMPORTANT
Design piping correctly, referring to the following to
prevent damage to sensors and to assure accurate
measuring.
NOTE
This section describes the remote sensor as an
example. The same attention must be paid to the
integral flowmeter.
(1) Location
IMPORTANT
Install the flowmeter in a location where it is not
exposed to direct sunlight. The minimum ambient
temperature is limited by the minimum fluid temperature
of the sensor (the lining). For more information, read the
applicable general specification as listed inTable 1.1.
The flowmeter may be used in an ambient humidity
where the relative humidity ranges from 0 to 100%.
However, avoid long-term continuous operation at
relative humidity 95% or higher.
(2) Noise Avoidance
IMPORTANT
The flowmeter should be installed away from electrical
motors, transformers, and other power sources in
order to avoid interference with measurement. When
installing two or more magnetic flowmeters, provide a
distance of at least 5D (D is size of model code) each
other. If diameters of them are different, let D be the
larger one.
Installation