![YOKOGAWA YTA Series Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/yokogawa/yta-series/yta-series_user-manual_3847593022.webp)
6-1
IM 01C50C05-01EN
6. Maintenance
6.1 General
Each component of this instrument is configured in units to make
maintenance easier.
This chapter contains trim adjustment and troubleshooting for maintenance
of the affected instrument.
IMPORTANT
Maintenance of this instrument should be performed in a service shop
where the necessary tools are provided.
6.2 Trim adjustment
This instrument is fully factory-tested and is guaranteed for the intended
accuracy, eliminating the need for the trim adjustment. When the trim
adjustment needs, the following procedure is recommended.
6.2.1 Trim Procedure
1. Trim sensor
The sensor trim procedure allows you to adjust for local conditions, changing
how the transmitter calculates process variables. There are two ways to
trim the sensor: Device variable zero trim and Device variable trim. Device
variable zero trim is a one-point adjustment typically used to compensate for
zero shifts. Device variable trim is a two-point process, in which two accurate
end-point sensor input are applied (equal to or greater than the range
values), and all output is linearized between them.
• Procedure to call up the D/A trim display
Device setup → Diag/Service → Calibration → Dev var zero trim
Device variable trim
In Device variable trim, if the output value is different from the trim setting