9-7
IM WT1801-01EN
Integration Conditions (Integ Set)
You can configure the following integration conditions.
•
•
Integration timer (Integ Timer)
•
Scheduled times for real-time integration (Real-time Control)
•
Integration auto calibration on/off (Auto Cal)
•
Watt hour integration method for each polarity (WP ± Type)
•
Current mode for current integration (q Mode)
•
Rated time of integrated D/A output (D/A Output Rated Time; option
Integration Mode (Mode)
The integration feature has the following five modes.
Integration mode
Start
Stop
Repetition
Manual integration mode
(Normal)
Key
operation
Key
operation
–––
Normal integration mode
(Normal)
Key
operation
Stopped by
the timer
–––
Continuous integration mode
(Continuous)
Key
operation
Key
operation
Repeats when
the timer expires
Real-time normal integration mode
(R
–
Normal)
Date and
time
Date and
time
–––
Real-time continuous integration mode
(R
–
Continuous)
Date and
time
Date and
time
Repeats when
the timer expires
Manual integration mode
When you set the integration mode to normal integration mode (Normal) and set the integration timer to
00000:00:00, the WT1800 performs integration in manual integration mode. After integration is started, it
continues until you press Stop. However, if either of the conditions below is met, integration is stopped, and the
integration time and integrated value are held.
• The integration time reaches the maximum integration time (10000 hours).
• The integrated value reaches its maximum or minimum displayable value.
Integration time
Reset
Start
Integrated value
Stop
Start
Reset
Hold
Hold
Hold
Hold
When the maximum
integrated value is
reached
9 Integrated Power (Watt hours)