8-9
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
6
Click the [Start Zero-point Adjustment] button. The mode transitions to the input
adjustment mode.
Adjustment takes up to 30 seconds.
When zero-point adjustment is completed, the following dialog is displayed.
7
Connect the calibration equipment to the channel for the analog input to adjust.
Input the reference value for the range to be adjusted according to the table below,
and click the [Start full-point adjustment] button.
Range
Reference value
20 mV
20 mV
60 mV
60 mV
200 mV
200 mV
1 V
1 V
2 V
2 V
6 V
6 V
20 V
20 V
60 V
60 V
200Ω
-3 wire
200Ω
400Ω
-3 wire
400Ω
200Ω
-4 wire
200Ω
400Ω
-4 wire
400Ω
2000Ω
-4 wire
2000Ω