![YOKOGAWA VigilantPlant PK200 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/yokogawa/vigilantplant-pk200/vigilantplant-pk200_user-manual_3499874021.webp)
<5.
Operation
>
5-3
IM 21B03D01-01E
Table 5.4
Output Signals
Output Signal Code
Output Signal
1
20 to 100 kPa
2
40 to 200 kPa
3
0.2 to 1.0 kgf/cm
2
4
0.4 to 2.0 kgf/cm
2
5
0.2 to 1.0 bar
6
0.4 to 2.0 bar
7
3 to 15 psi
8
6 to 30 psi
9
3 to 27 psi
(5) Input a 100% input signal.
Apply 20mA for 4 to 20mA input or 50mA for 10
to 50mA input.
Control the zero adjustment so that the output
air pressure reaches a specified pressure.
For 100% output air pressure, see the data
plate or Table 5.2 “Output Signals.” (See Figure
5.3.)
To increase output air pressure, please turn
span adjustment clockwise. To decrease output
air pressure, please turn span adjustment
counterclockwise. (See Figure 5.3.)
For the model with reverse action (/RA), please
turn span adjustment counterclockwise to
increase output air pressure. To decrease
output pressure, please turn span adjustment
clockwise.
(6) Input a 0% input signal.
Confirm if 0% output air pressure reaches the
specified pressure. If not, make zero point
adjustment.
(7) Input a 100% input signal
Confirm if 100% output air pressure reaches
the specified pressure. If not, make span
adjustment.
(8) Repeat steps (6) and (7) two to three times to
adjust the zero point and span.
(9) After completing adjustments, change the input
signal to 0%, 25%, 50%, 75%, and 100% and
confirm if the specified output is reached at the
relevant input.
Confirm the accuracy with reference to the input
and output signal reference values given in
Table 5.5 “Input and Output Signals.”
Zero adjustment
F0503.ai
Span adjustment
Figure 5.3
Span Adjustment Mechanism
Table 5.5
Input and Output Signals
Input Current of the model with / RA is shown in ( ).
Input
4 to 20mA
4mA (20mA)
8mA (16mA)
12mA (12mA)
16mA (8mA)
20mA (4mA)
10 to 50mA
10mA (50mA)
20mA (40mA)
30mA (30mA)
40mA (20mA)
50mA (10mA)
Output Signal
0 %
25 %
50 %
75 %
100 %
Output
Signal
1
20 kPa
40 kPa
60 kPa
80 kPa
100 kPa
2
40 kPa
80 kPa
120 kPa
160 kPa
200 kPa
3
0.2 kgf/cm
2
0.4 kgf/cm
2
0.6 kgf/cm
2
0.8 kgf/cm
2
1.0 kgf/cm
2
4
0.4 kgf/cm
2
0.8 kgf/cm
2
1.2 kgf/cm
2
1.6 kgf/cm
2
2.0 kgf/cm
2
5
0.2 bar
0.4 bar
0.6 bar
0.8 bar
1.0 bar
6
0.4 bar
0.8 bar
1.2 bar
1.6 bar
2.0 bar
7
3 psi
6 psi
9 psi
12 psi
15 psi
8
6 psi
12 psi
18 psi
24 psi
30 psi
9
3 psi
9 psi
15 psi
21 psi
27 psi
三千控制阀网