4-17
IM 04P01B01-17E
Commands
4
Example
Assign the 2-channel digital display to screen 04
and automatically switch the displayed channels
every 5 seconds.
VD 04,2CH digital,AUTO5S
For flag display
Syntax
VD p1,p2,p3
<terminator>
p1
Screen number (01 to 15)
p2
Display type
FLAG
Flag display (fixed)
p3
Display ON/OFF for each channel
Specify the ON/OFF state of each channel
with 1 or 0 (see the explanation).
Query
VD[ p1]?
Example 1 On the 6-dot model, assign the flag display to
screen 03 and display the flags of channel 01
and 02.
VD 03,FLAG,110000
Example 2 On the 2-pen model with the computation
function (/M1 option), assign the flag display to
screen 03 and display the flags of measurement
channel 02 and computation channels 0A, 0B,
0E, and 0F.
VD 03,FLAG,01-11001100
Description Use the format below to specify the display ON/
OFF state of each channel.
• List the values one after another for all
channels and separate the measurement
channels and computation channels with a
hyphen.
• Channels cannot be omitted.
• The characters after the hyphen are valid only
on models with the computation function (/M1
option).
For multiple display (display in which different
screens can be assigned to the top and bottom
sections
Syntax
VD
p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,p11,p12
<terminator>
p1
Screen number (01 to 15)
p2
Display type
Multiple display
Display in which different screens can
be assigned to the top and bottom
sections
p3
Top section display type
1CH digital
1-channel digital display
2CH digital
2-channel digital display
3CH digital
3-channel digital display (
µ
R20000)
TIME
Date/Time (
µ
R10000)
Chart speed
Chart speed (
µ
R10000)
TIME/Chart speed
Date/Time/Chart speed (
µ
R20000)
Channel alarm status
Channel alarm status display
STATUS
Status display
Lights out
Screen OFF
Tag_1CH digital
1-channel digital display (tag)
Note
Parameters that become invalid due to p3, p4, p8, or p9
setting are skipped, and the next parameter is brought
forward.
p4
Displayed channel switching interval
Condition: Set this parameter when p3 is
1CH digital
or
2CH digital
.
MANUAL
Fixed display channels, data
update rate (pen model): 2 s,
data update rate (dot model):
scan interval
AUTO1S
Switch the displayed
channels and data every 1
second.
AUTO2S
Switch the displayed
channels and data every 2
seconds.
AUTO3S
Switch the displayed
channels and data every 3
seconds.
AUTO4S
Switch the displayed
channels and data every 4
seconds.
AUTO5S
Switch the displayed
channels and data every 5
seconds.
p5
Channel number
Condition: This parameter can be specified
when p3 is
1CH digital
,
2CH digital
,
or
Tag_1CH digital
and p4 is
MANUAL
.
p6
2nd channel number
Condition: This parameter is specified when
p3 is
2CH digital
and p4 is
MANUAL
.
p7
3rd channel number
Condition: This parameter is specified when
p3 is
3CH digital
and p4 is
MANUAL
.
p8
Bottom section display type
Same as p3
p9
Displayed channel switching interval
Same as p4
p10
Channel number
Same as p5
p11
2nd channel number
Same as p6
p12
3rd channel number
Same as p7
4.4 Setting Commands