App-6
IM 04P01B01-17E
Appendix 4
A List of Error Messages
There are cases in which error codes and messages are displayed on the screen during
operation. The error messages and their description are listed below.
Setting Errors
Code
Message
Explanation/Countermeasures
1
System error.
Contact your nearest YOKOGAWA dealer.
2
Incorrect date or time setting.
Check the setting.
3
A disabled channel is selected.
The channel does not exist.
4
Incorrect function parameter.
Incorrect communication parameter.
5
The input numerical value exceeds the set range.
–
6
Incorrect input character string.
The entered character cannot be used.
7
Too many characters.
–
8
Incorrect input mode.
Incorrect range mode (Volt, TC, Scale, etc.) setting.
9
Incorrect input range code.
Incorrect range type (2 V, R, PT100, etc.) setting.
10
Format error.
Incorrect character string format.
11
Range settings are not same within the selected
Channels whose range differs cannot be set simultaneously.
channels.
12
An invalid characters.
Contains an invalid character.
13
Ref. CH error.
Specify a channel whose range is set to voltage, TC, or RTD for
the reference channel.
15
Too many characters for printout.
Too many characters in the message printout including
measured/computed values.
16
Too many characters for message.
Cannot enter a setting that exceeds the message string
(16 characters).
21
Cannot set an alarm for a SKIPPED channel.
–
22
The upper and lower span limits are equal.
This is not allowed.
23
The upper and lower scale limits are equal.
This is not allowed.
24
The lower span limit is greater than the upper span
–
limit.
25
The lower scale limit is greater than the upper scale –
limit.
26
Bias cannot be set to the SKIPPED channel.
–
27
Bias cannot be set to the DI channel.
–
30
The partial boundary value exceeds the range of the –
span.
31
Partial is invalid on the SKIPPED channel.
–
35
The upper and lower limits of the printing zone
Set the rightmost value of the zone – the leftmost value
≥
5 mm.
are equal.
36
The lower limit of the printing zone is greater
Set the rightmost value of the zone – the leftmost value
≥
5 mm.
than the upper limit.
37
The printing zone is narrower than the minimum
Set the rightmost value of the zone – the leftmost value
≥
5 mm.
width (5 mm).
38
Partial is invalid on the DI channel.
–
39
The bias and the calibration cannot be used
–
simultaneously.
40
Datume value(1 >= 2)
Set the measured value of the first point less than that of the
second point in the calibration correction.
41
Datume value(n-1 > n)
Set the measured value of the n–1
th
point less than or equal to
that of the n
th
point in the calibration correction.
42
Revise value(n-1 >= n)
Set the correction value of the n–1
th
point less than that of the n
th