<4. OPERATION>
4-12
IM 12T03A01-02E
4.4.3
Calibration Error
If an error indication (E005 in the first calibration or E006 in the second calibration) is displayed
when calibration computation finishes, examine the data input to function No. ‘3’, ‘4’, ‘5’, ‘7’, ‘8’ or
‘9’ and repeat the computation using the incorrect data.
4.5 Operation
4.5.1
Operation Procedure
Start the density measuring system, and monitor operating conditions for a while to confirm
that there is no abnormality. Adjustment for the density measuring system during operation is
unnecessary, however, daily check of liquid flow rate and pressure is preferable. When the steam
tracing pipe is installed in the sampling unit, do not heat the detector.
If an error indication is displayed, make a countermeasure immediately. Errors are listed in Table
4.2.
Table 4.2 Error Table
Display
Contents
Error Release and Remedy
E001
Abnormal detector density signal
When cause is removed
E002
Abnormal detector temperature sinal
When cause is removed
E003
Abnormal on converter
E004
Out of measuring temperature
E005
Abnormal first calibration
Check and reset the data required for
calibration.
E006
Abnormal second calibration
E007
Illegal data of EEPROM
Check and reset the loaded constant
NOTE
E001, 002, 003, 004 and 007 are cancelled temporally by pressing the FUNCTION No. key
in maintenance or calibration mode, however in the measuring mode, the error indication is
displayed again. These errors are detected only in the measuring mode.
E005 and E006 are errors that occur only in the calibration mode. You can clear these errors
by pressing the "SET" key. After that, the density at the reference temperature by the previous
calibrating data is displayed.
4.5.2
Check of Measured Density Value
It is recommended that measured density values are checked at intervals of several weeks to
determine the period for calibration and cleaning.
(1) Measure the density using a hydrometer to compare it with the density indication (at the
reference temperature) on the vibration type density meter. When sampling the liquid, check
the pressure and the temperature are not largely different from those of the normal state.
(2) Measure density correctly at the reference temperature which is same as the sample liquid
temperature.
(3) Compare the density measured at (2) with the displayed value when sampling the liquid. If
the difference of them exceeds the allowable error range, clean the vibrator.
For cleaning procedure, refer to Section 5.1.1.
4.5.3 Shutdown
When shutdown sample following steps and the vibration type density meter also stops, drain the
sample from the detector.
5th Edition : Dec. 11, 2015-00
Содержание vigilantplant DM8
Страница 1: ...User s Manual R Model DM8 Vibration Type Density Meter IM 12T03A01 02E IM 12T03A01 02E 5th Edition ...
Страница 5: ...Blank Page ...
Страница 9: ...Blank Page ...
Страница 25: ...Blank Page ...
Страница 41: ...Blank Page ...
Страница 57: ...Blank Page ...
Страница 63: ...Blank Page ...