IM 05P01F31-11EN  page 2/10

  Mounting the Instrument Main Unit

Provide an instrumented panel steel sheet of 1 to 10 mm thickness.

After opening the mounting hole on the panel, follow the procedures below to install 

the controller:

1)  Insert the controller into the opening from the front of the panel so that the 

terminal board on the rear is at the far side.

2)  Set the brackets in place on the top and bottom of the controller as shown in the 

figure below, then tighten the screws of the brackets. Take care not to overtighten 

them.

 

Bracket

(top mounting hardware)

Terminal board

Bracket

(bottom mounting hardware)

Panel

Insert a screwdriver into the

brackets to tighten the screws.

Direction to insert the

controller

Appropriate

tightening torque:

0.25 N

m

Insert the controller

into the opening at

the front of the panel.

CAUTION

•  Tighten the screws with appropriate tightening torque within 0.25 

N•m. Otherwise it may cause the case deformation or the bracket 

damage.

•  Make sure that foreign materials do not enter the inside of the

  instrument through the case’s slit holes.

  External Dimensions and Panel Cutout Dimensions

[(N-1)×48+45]

+0.6

0

65 (2.56)

48 (1.89)

11 (0.43)

UT32A

1 to 10 mm (0.04 to 0.39 inch) (panel thickness)

Bracket

Bracket Terminal cover

94.6 (3.72)

96 (3.78)

105.2 (4.14)

Unit: mm (approx. inch)

70 (2.76)

min.

(25)

145 (5.71)

min.

+0.6

0

45

(53) (2.09)

+0.8

0

92

• General mounting

• Side-by-side close mounting

20  (0.79)

91.6 (3.61)

“N” stands for the number of controllers to be

installed.

However, the measured value applies if N

5.

Normal tolerance: 

±(value of JIS B 0401-1998 tolerance class IT18)/2

+0.03

0

(3.62        )

+0.02

0

(1.77         ) (0.98)

+0.8

0

92

+0.03

0

(3.62        )

([(N-1)×1.89+1.77]        )

+0.02

0

4.  Hardware Specifications

WARNING 

This instrument is for Measurement Category No.1.

Do not use it for measurements in locations falling under Measure-

ment Categories No.2, No.3, and No.4.

Internal Wiring

Outlet

Entrance IV

Cable

III

T

O( I )

II

Category

IEC/EN/CSA

/

UL 

61010-1

EN 61010-2-030

Remarks

No.1

Measurement 

Category I

O (Other)

For measurements performed on circuits not direct

-

ly connected to MAINS.

No.2

Measurement 

Category II

Measurement 

Category II

For measurements performed on circuits directly 

connected to the low-voltage installation.

No.3

Measurement 

Category III

Measurement 

Category III

For measurements performed in the building instal-

lation.

No.4

Measurement 

Category IV

Measurement 

Category IV

For measurements performed at the source of the 

low-voltage installation.

■  Input Specifications

●Universal Input (Equipped as standard)

•  Number of inputs: 1

•  Input type, instrument range, and measurement accuracy: See the table below,

Input Type

 Instrument Range

Accuracy

ºC

ºF

Thermo-

couple

K

-270.0 to 1370.0ºC -450.0 to 2500.0ºF ±0.1% of instrument range ±1 digit for 

0°C or more

 

±0.2% of instrument range ±1 digit for 

less than 0°C 

 

±2% of instrument range ±1 digit for 

less than -200.0°C of thermocouple K 

±1% of instrument range ±1 digit for 

less than -200.0°C of thermocouple T

-270.0 to  1000.0ºC -450.0 to 2300.0ºF

-200.0 to  500.0ºC -200.0 to 1000.0ºF

J

-200.0 to 1200.0ºC -300.0 to 2300.0ºF

T

-270.0 to 400.0ºC

-450.0 to 750.0ºF

0.0 to  400.0ºC

-200.0 to 750.0ºF

B

0.0 to 1800.0ºC

32 to 3300ºF

±0.15% of instrument range ±1 digit 

for 400°C or more

 

±5% of instrument range ±1 digit for 

less than 400°C

S

0.0 to 1700.0ºC

32 to 3100ºF

±0.15% of instrument range ±1 digit 

R

0.0 to 1700.0ºC

32 to 3100ºF

N

-200.0 to 1300.0ºC -300.0 to 2400.0ºF

±0.1% of instrument range ±1 digit

±0.25% of instrument range ±1 digit 

for less than 0°C

E

-270.0 to 1000.0ºC -450.0 to 1800.0ºF ±0.1% of instrument range ±1 digit for 

0°C or more

 

±0.2% of instrument range ±1 digit for 

less than 0°C

 

±1.5% of instrument range ±1 digit for 

less than -200.0°C of thermocouple E.

L

-200.0 to 900.0ºC

-300.0 to 1600.0ºF

U

-200.0 to 400.0ºC

-300.0 to 750.0ºF

0.0 to 400.0ºC

-200.0 to 1000.0ºF

W

0.0 to 2300.0ºC

32 to 4200ºF

±0.2% of instrument range ±1 digit 

(Note 2)  

Platinel 2

0.0 to 1390.0ºC

32.0 to 2500.0ºF ±0.1% of instrument range ±1 digit  

PR20-40

0.0 to 1900.0ºC

32 to 3400ºF

±0.5% of instrument range ±1 digit for 

800°C or more 

 

Accuracy is not guaranteed for less 

than 800°C.  

W97Re3-

W75Re25

0.0 to 2000.0ºC

32 to 3600ºF

±0.2% of instrument range ±1 digit  

RTD

JPt100

-200.0 to 500.0ºC

-300.0 to 1000.0ºF ±0.1% of instrument range ±1 digit 

(Note 1) 

-150.00 to  150.00ºC -200.0 to 300.0ºF ±0.1% of instrument range ±1 digit

Pt100

-200.0 to 850.0ºC

-300.0 to 1560.0ºF ±0.1% of instrument range ±1 digit 

(Note 1) 

-200.0 to 500.0ºC

-300.0 to 1000.0ºF

-150.00 to 150.00ºC -200.0 to 300.0ºF ±0.1% of instrument range ±1 digit

Standard signal

0.400 to 2.000 V

±0.1% of instrument range ±1 digit

1.000 to 5.000 V

4.00 to 20.00 mA

DC voltage/current

0.000 to 2.000 V

0.00 to 10.00 V

0.00 to 20.00 mA

-10.00 to 20.00 mV

0.0 to 100.0 mV

The accuracy is that in the standard operating conditions: 23±2°C, 55±10%RH, and power 

frequency at 50/60 Hz.

Note 

1:  ±0.3°C ±1 digit in the range between 0 and 100°C, ±0.5°C ±1 digit in the range 

 

  between -100 and 200°C.

Note 2:  W: W-5% Re/W-26% Re(Hoskins Mfg.Co.). ASTM E988

• Input sampling (control) period: 200 ms

• Burnout detection:

    Functions at TC, RTD, and standard signal.

    Upscale, downscale, and off can be specified.

    For standard signal, burnout is determined to have occurred if it is 0.1 V or 0.4

 

  mA or less.

• Input bias current: 0.05 µA (for TC or RTD)

• Measured current (RTD): About 0.16 mA

• Input resistance:

    TC or mV input: 1 MΩ or more

    V input: About 1 MΩ

    mA input: About 250 Ω

• Allowable signal source resistance:

    TC or mV input: 250 Ω or less

   

Effects of signal source resistance: 0.1 µV/Ω or less

    DC voltage input: 2 kΩ or less

   

Effects of signal source resistance: About 0.01%/100 Ω

•  Allowable wiring resistance:

    RTD input: Max. 150 Ω/wire  (The conductor resistance between the three wires 

 

  shall be equal.)

    Wiring resistance effect: ±0.1ºC/10 Ω

•  Allowable input voltage/current:

    TC, mV, mA and RTD input: ±10 V DC

    V input: ±20 V DC

    mA input: ±40 mA

•  Noise rejection ratio:

    Normal mode: 40 dB or more (at 50/60 Hz)

    Common mode: 120 dB or more (at 50/60 Hz) 

    For 100-240 V AC, the power frequency can be set manually. Automatic

 

  detection is also available.

    For 24 V AC/DC, the power frequency can be set manually.

•  Reference junction compensation error:

    ±1.0ºC (15 to 35ºC), ±1.5ºC (-10 to 15ºC and 35 to 50ºC)

•  Applicable standards: JIS/IEC/DIN (ITS-90) for TC and RTD

■  Analog Output Specifications (Suffix code: -C)

•  Number of outputs:

    Control output: 1

•  Output type: Current output

•  Current output: 4 to 20 mA DC or 0 to 20 mA DC/load resistance of 600 Ω or less

•  Current output accuracy: ±0.1% of span (±5% of span for 1 mA or less)

    The accuracy is that in the standard operating conditions: 23±2°C, 55±10%RH, 

 

  and power frequency at 50/60 Hz.

■  Analog Output Specifications (Suffix code: -V)

•  Number of outputs:

    Control output: 1

•  Output type: Voltage pulse output

•  Voltage pulse output:

    Use: Time proportional output

    On-voltage: 12 V or more/load resistance of 600 Ω or more

    Off-voltage: 0.1 V DC or less

    Time resolution: 10 ms or 0.1% of output, whichever is larger

■  Relay Contact Output Specifications (Suffix code: R)

•  Contact type and number of outputs:

    Control output: contact point 1c; 1 point

•  Contact rating:

    Contact point 1c (control output): 250 V AC, 3 A or 30 V DC, 3A (resistance load)

•  Use: Time proportional output, ON/OFF output

•  Time resolution of control output: 10 ms or 0.1% of output, whichever is larger

Note:  The control output should always be used with a load of 10 mA or more.

■  Relay Contact Output Specifications

•  Contact type and number of outputs:

    Alarm output: contact point 1a; 2 points (common is independent)

•  Contact rating:

    Contact point 1a (alarm output): 240 V AC, 1A or 30 V DC, 1 A (resistance load)

•  Use: Alarm output, FAIL output, etc.

Note:  The alarm output should always be used with a load of 1 mA or more.

■  24 V DC Loop Power Supply Specifications

•  Use: Power is supplied to a 2-wire transmitter.

•  Power supply: 21.6 to 28.0 V DC

•  Rated current: 4 to 20 mA DC

•  Maximum supply current: About 30 mA (with short-circuit current limiting circuit.)

■  Safety and EMC Standards

•  Safety: 

  Compliant with IEC/EN 61010-1 (CE), IEC/EN 61010-2-201 (CE), IEC/EN 61010-

2-030 (CE), approved by CAN/CSA C22.2 No. 61010-1 (CSA), approved by UL 

61010-1.

  Installation category: II

  Pollution degree: 2

  Measurement category: I (CAT I) (UL, CSA)

  

 

    

O (Other) (CE)

  Rated measurement input voltage: Max. 10 V DC

  Rated transient overvoltage: 1500 V (*)

 

* This is a reference safety standard value for measurement category I of CSA/UL 61010-

1, and for measurement category O of IEC/EN 61010-2-030. This value is not necessarily a 

guarantee of instrument performance.

•  EMC standards:

  Compliant with CE marking

  EN 61326-1 Class A, Table 2 (For use in industrial locations),

  EN 61326-2-3

 

* The instrument continues to operate at a measurement accuracy of within ±20% of the 

range during testing.

  EN 55011 Class A, Group 1

  EN 61000-3-2 Class A

  EN 61000-3-3

  EMC Regulatory Arrangement in Australia and New Zealand (for all model includ-

ing LL50A)

 

EN 55011 Class A, Group 1

•  KC marking:

  Electromagnetic wave interference prevention standard, electromagnetic wave 

protection standard compliance

■  Construction, Installation, and Wiring

•  Dust-proof and drip-proof: IP66 (for front panel) (Not available for side-by-side 

close mounting.)

•  Material: Polycarbonate (Flame retardancy: UL94V-0)

•  Case color: White (Light gray) or Black (Light charcoal gray)

•  Weight: 0.5 kg or less

•  External dimensions (mm): 48 (W) × 96 (H) × 65 (depth from the panel face)

    (Depth except the projection on the rear panel)

•  Installation: Direct panel mounting; mounting bracket, one each for upper and 

lower mounting

•  Panel cutout dimensions (mm): 45

+0.6/0

 (W) × 92

+0.8/0

 (H)

• 

Mounting attitude: Up to 30 degrees above the horizontal. No downward titling allowed.

•  Wiring: M3 screw terminal with square washer (for signal wiring and power wiring)

■  Power Supply Specifications and Isolation 

•  Power supply:

    Rated voltage: 100-240 V AC (+10%/-15%), 50/60 Hz

                  

24 V AC/DC (+10%/-15%) (for /DC option)

•  Power consumption: 15 VA (DC:7 VA, AC: 11 VA if /DC option is specified)

•  Data backup: Nonvolatile memory

•  Power holdup time: 20 ms (for 100 V AC drive)

•  Withstanding voltage

 

Between primary terminals and secondary terminals: 2300 V AC for 1 minute (UL, CSA)

  Between primary terminals and secondary terminals: 3000 V AC for 1 minute (CE)

    Between primary terminals: 1500 V AC for 1 minute

    Between secondary terminals: 500 V AC for 1 minute

          (Primary terminals: Power* and relay output terminals; Secondary terminals: 

 

        Analog  I/O  signal  terminals,  communication  terminals  and  functional 

 

        grounding terminals.)

 

*: Power terminals for 24V AC/DC models are the secondary terminals.

•  Insulation resistance: Between power supply terminals and a grounding terminal 

20 MΩ or more at 500 V DC

•  Isolation specifications

The circuits divided by lines are insulated mutually.

Control (voltage pluse, analog) output terminal 

Control relay (c-contact) output terminal

PV (universal) input terminal

Alarm-1 relay (a-contact) output terminal

Alarm-2 relay (a-contact) output terminal

Internal 

circuits

Power 

supply

24 V DC loop power supply terminal

RS485 communication terminal

■  Environmental Conditions

Normal Operating Conditions:

•  Ambient temperature: -10 to 50°C (-10 to 40°C for side-by-side mounting of con-

trollers)

•  Ambient humidity: 20 to 90% RH (no condensation allowed)

•  Magnetic field: 400 A/m or less

•  Continuous vibration at 5 to 9 Hz: Half amplitude of 1.5 mm or less, 1oct/min for 90 

minutes each in the three axis directions

  Continuous vibration at 9 to 150 Hz: 4.9 m/s

2

 or less, 1oct/min for 90 minutes each 

in the three axis directions

•  Short-period vibration: 14.7 m/s

2

, 15 seconds or less

•  Shock: 98 m/s

2

 or less, 11 ms

•  Altitude: 2000 m or less above sea level

•  Warm-up time: 30 minutes or more after the power is turned on

•  Startup time: Within 10 seconds

 

*:  The LCD (a liquid crystal display) is used for a display portion of this product.

 

  The LCD has a characteristic that the display action becomes late at the low 

 

  temperature. However, the control function is not affected. 

Transportation and Storage Conditions:

•  Temperature: -25 to 70ºC

•  Temperature change rate: 20ºC/h or less

•  Humidity: 5 to 95% RH (no condensation allowed)

Effects of Operating Conditions

•  Effect of ambient temperature:

    Voltage or TC input: ±1 µV/ºC or ±0.01% of F.S./ºC, whichever is larger

    Current input:  ±0.01% of F.S./ºC

    RTD input: ±0.05ºC/ºC (ambient temperature) or less

    Analog output: ±0.02% of F.S./ºC or less

•  Effect of power supply voltage fluctuation

    Analog input: ±0.05% of F.S. or less

    Analog output: ±0.05% of F.S. or less

     (Each within rated voltage range)

Содержание UT32A-R

Страница 1: ...umented panel Mount the instrument in a location where its terminals will not inadvertently be touched Well ventilated locations Mount the instrument in well ventilated locations to prevent the instru...

Страница 2: ...e set manually Reference junction compensation error 1 0 C 15 to 35 C 1 5 C 10 to 15 C and 35 to 50 C Applicable standards JIS IEC DIN ITS 90 for TC and RTD Analog Output Specifications Suffix code C...

Страница 3: ...urrent as a spark removal surge suppression circuit into the line in parallel with the load After completing the wiring the terminal cover is recommended to use for the instrument Recommended Crimp on...

Страница 4: ...IM 05P01F31 11EN page 4 10 Intentionally blank Intentionally blank...

Страница 5: ...to easily set the basic function of the controller Turn on the controller to start the Quick setting function This function allows you to easily set the control type and input and quickly start the c...

Страница 6: ...nt Open unlit SP or Target SP Alarm setpoint Hysteresis SP or Target SP Hysteresis Alarm setpoint Alarm setpoint Output Hysteresis Closed lit Hysteresis Alarm setpoint Output Alarm setpoint Hysteresis...

Страница 7: ...arrow keys to display the required setpoint Blinks during the change The setpoint has been registered This starts auto tuning The limiter can be set to the output during auto tuning For details see U...

Страница 8: ...Contact us for repair ERR SYS System data error System data is corrupted Faulty Contact us for repair PAR 0004 for user default value error only Setup parameter PA ER User parameter default value erro...

Страница 9: ...13 Target SP deviation high and low limits 14 Target SP deviation within high and low limits 15 OUT high limit 16 OUT low limit 19 Analog input PV high limit 20 Analog input PV low limit 29 PV velocit...

Страница 10: ...Ladder communication CO M Coordinated master station CO S Coordinated slave station MBASC Modbus ASCII MBRTU Modbus RTU CO S1 Coordinated slave station Loop 1 mode CO S2 Coordinated slave station Loo...

Отзывы: