Manual Change for IM 01G05B02-01E 1st Edition : No.15-021-E Nov. 24 2015-00
Addition to the section
“4.2.1 Key Operations” (page 16) and related pages :
Refer to the following table for the function of corresponding key operations.
The correct explanations for these key operations in the related pages are also as the table below.
Related pages:
“16.2 US300PM doesn’t react anymore” (page 113)
RESET
INIT
BRK
ENTER
[Operation during the measurement or menu display]
RESET: Press these three keys simultaneously to recover from an error. This has
the same effect as restarting the unit. Data will not be affected.
INIT (cold start): Press these three keys simultaneously and release
ENTER
first.
After acknowledging the display of main menu, release (i)
BRK
key first
and then (ii)
C
key. This will initialize the instrument. Most parameters and
settings are reset to the factory default values. The memory will not be
cleared.
Note that when the data logging function is activated, the
DELETE
MEAS.VAL.
display will appear instead of main menu. In this case, after
releasing (i)
BRK
key first and then (ii)
C
key, select
NO
or
YES
and then
press
ENTER
to finish the procedure.
ON
INIT
BRK
[Operation when powering the instrument on]
INIT (coldstart): Pressing these two keys simultaneously and after acknowledging the display
of main menu, release (i)
BRK
first and then (ii)
C
. This will initialize the instrument.
Most parameters and settings are reset to the factory default values.
The same procedure is required as above when the
DELETE MEAS.VAL.
display
appears.
Additio
n to the section “5.6 Selection of the Sound Path Factor” (page 30) :
When you select “auto” for “Sound Path” described at the section “11.2 Settings for Dialogues and Menus”, the
setting for the sound path factor will be automatic. In this case, “Reflection Mod” display will appear and you
can select either “Diagonal Mode” by “no” or “Reflection Mode” by “yes”.
A:Reflection Mod
no >YES<
When the pipe outer diameter is more than 600m
m, the sound path factor as “1” (one) is recommended.
Otherwise, the measurement may become unstable when flow velocity or fluid temperature changed.
Addition to the section “5.7.2 Mounting of the Transducers” (page 31) :
The pipe wall thickness may slightly vary from part to part. Check it in advance by applying a wall thickness
probe or some other ways and avoid mounting the transducers on such parts.
Addition to the section “5.7.2.4 Mounting with Fixtures” (page 32) :
Attention!
When using mounting fixtures, there may arise some air gap between the transducer
surface and pipe wall because of any distortion of the pipe wall. Make sure to avoid
having such air gap between them.