background image

43

3.  Specifications

3.1  General Specifications

Measurement function:
 

DC Voltage, AC Voltage, DC Current, AC Current, Resistance, 
Frequency, Temperature, Capacitor, Continuity Check, Diode Test

Other functions: 

Data Hold (D•H), Auto Hold (A•H), Range Hold (R•H), Maximum 

value* (MAX), Minimum value* (MIN), Average value* (AVG), 
Zero Adjustment (Capacitor, Resistance), Relative values, Save to 
Memory*, LCD backlight.

 

*: For model TY530 only

Measuring method: 

Σ modulation

Display: 

4-digit (LCD)/7-segment 

 

Maximum Reading: 6000

 

Polarity Indicator: “–” Appears automatically  when the polarity is 

negative

 

Overrange Indicator: “ OL ”

 

Low-battery Indicator: “

 ” Appears when the batteries become 

low

Measurement cycle:   5 times per second
 

(except frequency measurement : one time per second, Resistance 

measurement  (6MΩ/60MΩ)  :  2.5  times  per  second,  capacitor 
measurement (1000μF) : max.0.14 time per second)

 

Bar graph display approx 25 times per second (at AC, Ω)

Operating temperature and humidity ranges:

 

-10 to 55ºC, 80%RH or less (no condensation)

 

70%RH or less at 40 to 55ºC.

Storage temperature and humidity ranges:

 

-30 to 70ºC, 70%RH or less (no condensation)

Temperature coefficient:

 

(Accuracy at 23±5ºC× 0.1)/ºC should be added.

 

(Temperature ranges: -10 to 18ºC and 28 to 55ºC)

Power supply: 

AA-size (R6/LR6) 1.5V batteries: 4

Battery life: 

Approximately 300 hours 

 

(Operating hours of alkaline batteries when in DC voltage-mode.)

 

Note: The battery life varies depending on the operating conditions.

Insulation resistance : 1000V DC, 100MΩ or more

Содержание TY520

Страница 1: ...IM TY530 1st Edition Nov 2008 KYOU Japanese English Store this manual in a safe place for future reference TY520 TY530 Digital Multimeter User s Manual...

Страница 2: ...loss of life Yokogawa is in no way liable for any damage resulting from the user s mishandling of the product For safe use of this product the following safety symbols are used on the product n About...

Страница 3: ...ion that is essential for handling the instrument or should be noted in order to familiarize yourself with the instrument s operating procedures and or functions Danger Handle with Care This symbol in...

Страница 4: ...sting leads to from the instrument Disconnect testing leads from the instrument before opening the casing to replace the batteries or for any other reason A cap is provided on the tip of a testing lea...

Страница 5: ...ment 57 4 3 9 Capacitor Measurement 57 4 3 10 Frequency Measurement 58 4 3 11 Function to change RMS detection to from MEAN detection mode TY530 only 58 4 3 12 Function to turn the filter on off 59 4...

Страница 6: ...t mean square value detection RMS and mean value detection MEAN can be switched during AC voltage measurement TY530 only l Low pass filter The low pass filter can be switched on off during AC voltage...

Страница 7: ...to the low voltage installation Appliances portable equipment ect CAT III For measurements performed in the building installation Distribution board circuit breaker ect CAT For measurements performed...

Страница 8: ...eries become low Measurement cycle 5 times per second except frequency measurement one time per second Resistance measurement 6M 60M 2 5 times per second capacitor measurement 1000 F max 0 14 time per...

Страница 9: ...gnetic field of 3 V m accuracy is within five times the rated accuracy Standard accessories Batteries 4 Testing leads 1set Model 98015 Fuse included 440mA 1000V 99015 10A 1000V 99016 User s manual 1 B...

Страница 10: ...e Resolution Accuracy Input Impedance Maximum Input Voltage 50 60Hz 40Hz to 500Hz 500Hz to 1kHz 600mV 0 1mV 0 5 5 1 5 1 5 5 10M 200pF 1000V rms AC 1000V DC 6V 0 001V 11M 50pF 60V 0 01V 10M 50pF 600V 0...

Страница 11: ...max Resistance Measurement Range Resolution Accuracy Maximum Measuring Current Open loop Voltage Input Protective Voltage 600 0 1 0 4 1 1 2mA 3 5V 1000V rms 6k 0 001k 110 A 1 3V 60k 0 01k 13 A 600k 0...

Страница 12: ...tive Voltage 10nF 0 01nF 2 10 1000V rms 100nF 0 1nF 2 5 1 F 0 001 F 10 F 0 01 F 100 F 0 1 F 3 5 1000 F 1 F Accuracy is specified after zero adjustment at 10n to 1 F Capacitance Frequency Measurement H...

Страница 13: ...ng Don t remove the Blank cover except when the USB adapter or Printer adapter is connected TY530 only Do not use the instrument near noise emitting equipment or where there may be a sudden change of...

Страница 14: ...nction switch Input terminal block SHIFT key RANGE key HOLD SAVE key LIGHT READ key RE SELECT PRINT key L key MIN MAX LOG key TY530 only MEMORY key TY530 only Black Red Caps of Testing leads Actions i...

Страница 15: ...nt mV AC voltage mV measurement AC SENSOR measurement in the SENSOR mode Diode Test A mA A AC current measurement 3 RANGE key Allows the operator to select the measuring range Fixed ranges The display...

Страница 16: ...ol The sub display shows the reference value 7 MIN MAX key TY530 only Displays the minimum value MIN maximum value MAX and average value AVG during measurement Pressing this key starts recording and a...

Страница 17: ...for resistance measurement Unit for temperature measurement Hz kHz Unit for frequency measurement Unit for percentage calculation mV Sub display Unit for SENSOR mode measurement Input voltage s Sub d...

Страница 18: ...surement V mV 1 Turn the function switch to the V or mV position 2 Press the SELECT key when selecting the mV is displayed 3 Plug the testing leads into the input terminals 4 Connect the testing leads...

Страница 19: ...t clamp sensor The sub display shows the output voltage of the current clamp sensor mV input voltage of the instrument mV The main display shows the current values based on the preset setting of conve...

Страница 20: ...avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument 1 Turn the function...

Страница 21: ...ature Measurement TEMP CAUTION To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to th...

Страница 22: ...more than 440mA before the A or mA position is selected 2 Please select between DC and AC When selecting AC press the SELECT key 3 Plug the black testing lead into the COM input terminal and the red...

Страница 23: ...being applied to the instrument 1 Turn the function switch to the V position 2 Press the SELECT key to select the range of frequency The display shows the unit of frequency 3 Plug the testing leads i...

Страница 24: ...rn the filter on off during AC measurement 1 Select the appropriate AC measurement mode ACV ACmV AC A ACmA ACA by using the function switch and the SELECT key 2 Press the SHIFT key to display Shift on...

Страница 25: ...y measurement The Auto hold function can not be applied to unstable signals 4 3 14 Relative and percentage calculation The instrument can calculate relative values or difference and percentage values...

Страница 26: ...in or more steps of 1 min Press the HOLD key to stop recording The display shows the D H symbol To confirm the recording time For confirming the recording time press the MIN MAX key Subsequent pressin...

Страница 27: ...nt allocates the smallest number between 0000 to 1599 that has not yet been used Use the RANGE key or REL key switches the number of saved data To save a Data SAVE mode 1 Press the MEMORY key The disp...

Страница 28: ...LD mode disables HOLD mode To load a Data SAVE mode 1 Press the MEMORY key The display shows the MEM symbol 2 Press the READ LIGHT key 3 Press the SAVE HOLD key to select the number of saved data Sele...

Страница 29: ...deleted MIN MAX LOG 4 5 AUTO POWER OFF Function To use the AUTO POWER OFF function The display shows the AUTO OFF indication The instrument automatically turns off twenty minutes after the last key op...

Страница 30: ...preset mode 1 Press the SHIFT key shows Shift on the display 2 Press the LIGHT key changes the mode to Set up mode from Set up to AC detection method TY530 beep on off of buzzer TY520 TY530 TY520 3 Pr...

Страница 31: ...aving interval by using the RANGE key or REL key 3 Press the HOLD key to save the setting SEt appears and then L Int Settings of saving interval 1 2 5 10 30 60 600 1800 sec Sound on off setting Set th...

Страница 32: ...the SHIFT key switches them in the reversed sequence 2 Select each number place of decimal point and unit by using the RANGE key or REL key 3 Press the HOLD key to save the setting SEt appears and the...

Страница 33: ...he desired position of any measurement mode POWER ON state This enables the following functions corresponding to the press keys Keys Functions to be set SELECT LCD check Lit only while pressing the SE...

Страница 34: ...esting leads Before carrying out calibration read the instruction manual of the calibrator Temporarily remove the testing leads from the instrument before switching measurement mode function Condition...

Страница 35: ...e Carry out calibration of other ranges by repairing steps 6 and 7 with reference to Table 1 9 To quit calibration turn the function switch back to the OFF position Note Calibration should be started...

Страница 36: ...to OFF turn off the power Do not operate the instrument with the casing left open CAUTION Do not mix batteries of different types or new batteries with used ones Make sure the polarities of the new ba...

Страница 37: ...function switch to OFF turn off the power Do not operate the instrument with the casing left open In order to avoid damage to the instrument or any possible accident use fuses of the specified rating...

Страница 38: ...ical and Electronic Equipment WEEE Directive 2002 96 EC This Product complies with the WEEE Directive 2002 96 EC marking requirement The affixed product label see below indicates that you must not dis...

Страница 39: ...t of the People s Republic of China They are applicable only in the People s Republic of China Pb Hg Cd Cr VI PBB PBDE ASSY 98015 SJ T11363 2006 SJ T11363 2006 90050 90056 TC K 2006 2 28 SJ T11364 200...

Страница 40: ...Phone 55 11 5681 2400 Facsimile 55 11 5681 4434 YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION KOREA Phone 82 2 551 0660 to 0664 Facsimile 82 2 551 0665 YOKOGAWA AUSTRALIA PTY LTD AUSTRALIA Phone 6...

Отзывы: