To save a Data (LOGGING-mode)
1) Press the MEMORY key. (The display shows the “ MEM ”symbol.)
2) Press the LOG (MIN/MAX) key.
(The sub-display shows the logging interval (period).)
Set the value with the
(RANGE) key or
(REL
/
%) key.
The default setting is one second.
(The default settings can be changed. Refer to the Set-up function.)
The display shows “ FULL ” when the logging data is already saved.
When saving the new data, perform delete of data.
3) Press the LOG (MIN/MAX) key to start logging. (The “ MEM ”symbol is flashing.)
4) To cancel the function, hold down the MEMORY key for one second.
When memory capacity becomes full, the function is automatically canceled.
(“ MEM ” symbol disappears.)
Note
LOGGING-mode operation during HOLD-mode disables HOLD-mode.
To load a Data (SAVE-mode)
1) Press the MEMORY key. (The display shows the “ MEM ”symbol.)
2) Press the READ (LIGHT) key.
3) Press the SAVE (HOLD) key to select the number of saved data.
Select the number with the
(RANGE) key or
(REL
/%) key.
4) To cancel the function, hold down the MEMORY key for one second.
(“ MEM ” symbol disappears.)
To load a Data (LOGGING-mode)
1) Press the MEMORY key. (The display shows the “ MEM ”symbol.)
2) Press the READ (LIGHT) key.
3) Press the LOG (MIN/MAX) key to select the number of saved data.
Select the number with the
(RANGE) key or
(REL
/
%) key.
4) To cancel the function, hold down the MEMORY key for one second.
(“ MEM ” symbol disappears.)
<To delete of saving data>
Delete method (SAVE-mode)
• To delete all data
1) Press the MEMORY key.
(The display shows the “ MEM ”symbol.)
2) Hold down the SAVE (HOLD) key for one second.
(The display shows the “ CLr ? ”symbol.)
3) Press the SAVE (HOLD) key.
All data is deleted.