Quick Reference Instruction Manual
Wiring
Transmitter
46 / 54
IM01U10A00-00EN-R, 2nd edition, 2017-08-29
Status input
Sin
Do not connect a signal source with electric voltage.
The status input is provided for use of voltage-free contacts with the following specifica-
tion:
Switching status
Resistance
Closed
< 200 Ω
Open
> 100 kΩ
ROTAMASS
Sin+
Sin-
Fig. 38:
Status input connection
7.4.7 Connecting power supply and external devices
NOTICE
Risk of damage to the flow meter due to incorrect power supply
▶
The specified power supply must be observed (see General Specifications).
▶
The power-supply cable must be designed for the power supply used with a
minimum diameter of 0.5 mm.
2
1
1
Transmitter back cover
2
Power supply cable gland