IM 12D08N05-01E
5-5
<5. MENU STRUCTURE COMMISSIONING>
Concentration measurement is only possible if “measure” in the
“Configure sensor” menu is set to “conduc concentration”
or “Concentration only”.
Enter
Measurement setup
Temperature settings
Temp. Compensation
Calibration settings
Concentration
Measure Conductivity only
Configure sensor
Menu
Parameter
Default values
Range
min.
max.
Calibration
Air adjust
10.00 μS
0 μS
20 μS
c.c. high
120%
100%
120%
c.c. low
80%
80%
100%
Stabilization time
5 s
2 s
30 s
Calib. interval
250 days
1 day
250 days
Concentr. Table
Table
See appendix
Содержание SC450G
Страница 1: ...User s Manual IM 12D08N05 01E 5th Edition Model SC450G Conductivity Converter Style S2 ...
Страница 12: ...Blank Page ...
Страница 58: ...Blank Page ...
Страница 62: ...Blank Page ...
Страница 79: ...IM 12D08N05 01E APP 13 APPENDIXES Appendix 7 Control drawing for FM approval ...
Страница 80: ...Blank Page ...