IM 12A06S01-01EN-P
9
4-5. Model and suffix code
Model
Suffix
Option Description
code code code
SA11
SENCOM Smart Adapter
Measuring
-P1
pH/ORP, conventional
Parameter
-P2
pH/ORP, differential
-C1
Specific conductivity (SC)
Type
-AA
General purpose
Region
-N
Not specified
Connection type -VS
Variopin connector for SENCOM ID-chip in sensor
Style
-NN
Always -NN
Option
/UM
Pipe and wall mounting hardware
4-6. Spare part list
Order no.
Product Name
Quantity
Remark
K1548PQ
Universal mounting set
1 set
Option /UM
K1548GF
O-ring set (5 pieces)
1 set
Option /UM contains Pipe and Wall mounting hardware
Figure 4.1 Positioning of O-ring
5. Wiring and installation
Figure 5.1: Installation diagram SA11
5-1. Installation site
The SA11 is weatherproof and can be installed both inside and outside. Select an
installation site that meets the following conditions:
• Mechanical vibrations and shocks are negligible
• No relay switch and power switch are installed close to the device
• There is space mounting the device to the Yokogawa- sensor & Yokogawa HOST system
• Not exposed to direct sunlight or severe weather conditions
• Maintenance is possible
• No corrosive atmosphere
• Water Protection: IP67, NEMA Type 4X
YPA Europe
Made in the Netherlands 3825HD-2
BA11
ACTIVE JUNCTION BOX
BUS
SENSOR CONNECTIONS
IN
OUT
J1
J2
J3
J4
WU11-M9-**-CN-S
Modbus Host
e.g. FLXA402
WU11-M9-**-WP-S
WU11-M9-**-WP-S
max. 100m (328ft)
max. 100m
(328ft)
max. 100m
(328ft)
Suffix
-C-AA-M9
Serial#
N3T400171
Power
+2.7. +5.5VDC/30mW
T amb.
-20.. + 55ºC
IP
Type 4X / IP66
2018
Read IM
12B06W03-01E-E
Modbus Slave
SA11
Modbus Slave
SA11