IM 12A06S01-01EN-P

11

5-3.  Wiring to the HOST system

Wired connection 

The YOKOGAWA  HOST system is connected to the SA11 using the WU11 type S 

interconnection cable with at one end wired terminals and on the other end a M9 connector. 

This cable is specified for reliable transfer of digital signals, and especially designed to 

be installed in a heavy industrial environment. The double shielded cable will protect the 

connected devices for interference from high voltages and currents which are present on other 

cables. The cable is specified to be used in IP67 applications, allowing it to be submerged as 

a whole. The WU11 type S cable is available length up to a maximum of 100 meter (328ft).

Connection concept

The YOKOGAWA HOST system model FLXA402 can be connected to maximum five SA11 

devices. In case of more than two SA11 devices connected, the BA11 Active Junction Box 

must be used. Connection in between the FLXA402 and BA11 is done by using the WU11 

type -S interconnection cable with at one end wired terminals and on the other end a M9 

connector. Connection in between the BA11 and the SA11 is done using the WU11 type -S 

interconnection cable with at both ends a M9 connector. In both cases the maximum length 

has to be within 100 meter (328ft).

Wireless connection

In laboratory environments the HOST Fieldmate system can be connected to the SA11 

using the YOKOGAWA Bluetooth device model IB100. Interconnection in between the 

IB100 and SA11 is done using the WU11 type -S interconnection cable with at both ends a 

M9 connector. The wireless communication in between the Fieldmate system and IB100 is 

limited to a distance of 10 meter. 

Grounding

The measuring system must always be 

connected to a class D ground (a ground 

resistance of 100 Ohm or less), preferably 

done at the HOST side. To protect the 

SA11 device against interference, such as 

electromagnetic noise or common mode, all 

wires in WU11 type -S cable are connected 

at HOST. This makes the housing of the 

SA11 grounded. For correct measurement 

without loss of specifications, it is strongly 

recommended to ensure that the process 

liquid is grounded. Figure 5.4 shows the 

connection diagram in case without/UM 

option.

However, in case that there is different potential between Ground at HOST side and Ground 

which the housing of the SA11 is connected, figure 5.5 shows the example of the installation. 

To prevent ground loop by this different potentials it is necessary to disconnect one of the 

ground references. This can be done at the HOST side by disconnecting the wire #82 of the 

WU11 type S cable. Please connect #82 to NC terminal on the SENCOM SA module when 

SA11 is connected to FLXA402. Please be sure to insulate contact pin of #82 wire from 

terminals and metal housing, when SA11 is connected to other host. The SA11 device must 

then be installed using the optional pipe and wall mounting hardware (/UM) with a ground wire 

Host

Ground 1

Sensor

SA11

WU11

(max 100m)

Liquid

Ground 2

Figure 5.4: Typical installation of  

SA11 to host

Содержание SA11

Страница 1: ...IM 12A06S01 01EN P 2nd edition Start up Manual Model SA11 Smart Adapter...

Страница 2: ...tices 6 2 7 SA11 fonts 6 3 CE marking products 7 3 1 Authorized Representative in EEA 7 3 2 Identification Tag 7 3 3 Users 7 4 Introduction and general description 7 4 1 Instrument check 7 4 2 Checkin...

Страница 3: ...IM 01R01A01 01E Available for download The E or EN in the document number is the language code for English version the JA indicates the Japanese version and the Z indicates it is a multi language docu...

Страница 4: ...anual WARNING This symbol indicates that an operator must follow the instructions laid out in this manual in order to avoid the risks for the human body of injury electric shock or fatalities The manu...

Страница 5: ...od of typically 12 months from the date of shipment from the manufacturer Individual sales organizations can deviate from the typical warranty period and the conditions of sale relating to the origina...

Страница 6: ...fonts c Copyright 2000 2001 efont The Electronic Font Open Laboratory All rights reserved Redistribution and use in source and binary forms with or without modification are permitted provided that th...

Страница 7: ...se SA11 with YOKOGAWA s HOST systems and sensors Please read carefully this manual and the instruction manual of the relevant HOST system and sensor before using this SA11 slave device 4 1 Instrument...

Страница 8: ...cuits in which measures are taken to limit transient over voltages to an appropriately low level Note 2 Pollution degree indicates the degree of existence of solid liquid gas or other inclusions which...

Страница 9: ...ram SA11 5 1 Installation site The SA11 is weatherproof and can be installed both inside and outside Select an installation site that meets the following conditions Mechanical vibrations and shocks ar...

Страница 10: ...the SA11 on top of the YOKOGAWA labelled sensor or when the process conditions are higher than 100 C for power supply 2 7 to 4 5VDC or 125 C for power supply 4 5 to 5 5VDC an alternative mounting met...

Страница 11: ...IB100 Interconnection in between the IB100 and SA11 is done using the WU11 type S interconnection cable with at both ends a M9 connector The wireless communication in between the Fieldmate system and...

Страница 12: ...r s Manual If the system is configured incorrectly an error indicator may be displayed or the measurement values displayed may be incorrect Refer to the User Manual of the HOST system how to check the...

Отзывы: