IM MY600-01EN
2.
Set the range switch to the V/Ω position.
3.
Connect the other end of the probe (black) connected
to the EARTH terminal to the earth side of the circuit
under test. Connect the other end of the probe (red)
connected to the LINE terminal to the line side.
Select the V/Ω range.
98008
98009
Red
Black
4.
Check the reading without pressing the measurement
switch or remote switch. The instrument detects and
indicates DC or AC automatically.
• When a negative DC voltage is detected at the line
probe side, a minus sign appears to the left of the
displayed voltage.
• The instrument cannot detect DC or AC if the
measured value is less than 2 V.
Note
Measured results exceeding the display range (over-range)
are shown as follows:
AC voltage: >629 V
Positive DC voltage: >629 V
Negative DC voltage:<-629 V