46
IM MY600-01EN
Pass/Fail Judgment
Reference Value for Pass/Fail Judgment
For insulation resistance measurements, this instrument
has a judgment function that compares measurements
to the reference value and indicates the results using the
backlight and buzzer. This function can be turned off. You
can set the reference value to any value of your choice.
The result of a pass/fail judgment is indicated using the
backlight as follows:
Result of judgment
Backlight color
Display
Greater than the
reference value
Green
PASS
Less than or equal to the
reference value
Red
FAIL
When the measured value is
greater than the reference
value, “PASS” is displayed,
and the backlight blinks
green.
When the measured value
is less than or equal to the
reference value, “FAIL” is
displayed, and the
backlight blinks red.