6.6 General Specifications
Item
Specifications
Safety standards
1
Compliant standards
EN 61010-1
Overvoltage Category II
2
Pollution degree 2
3
EN 61010-2-030
Measurement Category Other (O)
4
Emissions
1
Compliant standards
EN 61326-1 Class A
EN 55011 Class A, Group1
EN 61000-3-2
EN 61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard (
한국 전자파적합성기준
)
This product is classified as Class A (for use in industrial environments). Operation of
this product in a residential area may cause radio interference, in which case the user
will be required to correct the interference.
Cable conditions
USB port (USB)
Use a shielded cable that is 3 m in length or less.
GP-IB connector (GP-IB)
Use a shielded cable that is 3 m in length or less.
Ethernet port (ETHERNET)
Use a category 5 or better Ethernet cable (STP) that is 30 m in length or less.
Trigger/sync input input terminal (TRIG IN/SYNC IN)
Use a BNC cable that is 3 m in length or less.
Sync output terminal (SYNC OUT)
Use a BNC cable that is 3 m in length or less.
DMM/24VOUT terminal (on models with the /DM option)
Use the included measurement lead (366961).
D/A output terminal (on models with the /DA option)
Use a BNC cable that is 3 m in length or less.
Comparator output terminal (on models with the /DA option)
Use the included terminal plug (A2119JT) and a cable that is 3 m in length or less.
Immunity
1
Compliant standards
EN 61326-1 Table 2 (for use in industrial locations)
Influence in the immunity test environment: Within ±20% of the range setting
However, the measured values may receive noise effects.
Cable conditions
Same as the emission cable conditions.
Environmental
standards
1
Compliant standards
EN 50581 Monitoring and control instruments including industrial monitoring and control
instruments.
Pressure directive
1
Sound Engineering Practice (SEP) as defined in the Pressure Equipment Directive (PED)
1 Applies to products with CE marks. For information on products without CE marks, contact your
nearest YOKOGAWA dealer.
2 The overvoltage category (installation category) is a value used to define the transient overvoltage
condition and includes the rated impulse withstand voltage. The overvoltage category II applies to
electrical equipment that is powered through a fixed installation, such as a switchboard.
3 Pollution Degree applies to the degree of adhesion of a solid, liquid, or gas that deteriorates
withstand voltage or surface resistivity. Pollution degree 2 applies to normal indoor atmospheres (with
only non-conductive pollution).
4 The measurement category of the instrument signal terminals is Other (O). Do not use it to measure
the main power supply or for Measurement Categories II, III, or IV.
Measurement category O (Other) applies to measurement of circuits that are not directly connected
to a main power supply. This category applies to measurement of secondary electric circuits in
equipment across a transformer. The estimated transient overvoltage that may appear at the
instrument’s signal input is 350 V.
Measurement category II applies to measurement of circuits, such as household electric appliances
and portable electric tools, that are connected to low-voltage installations.
Measurement category III applies to measurement of facility circuits, such as distribution boards
and circuit breakers.
Measurement category IV applies to measurement of power source circuits, such as entrance
cables to buildings and cable systems, for low-voltage installations.