5-13
IM MT300-02EN
Troubleshooting, Maintenance, and Inspection
1
2
3
4
5
6
App
Index
5.8 Adjustment and Calibration
Procedure
If you are calibrating the instrument, enter the calibration date.
Entering the User Calibration Date
1.
Press
MENU
to display the top menu.
2.
Press the
Utility
soft key,
the Next
soft key, and then the
Next
soft key. The following menu
appears.
3.
Press the
Maintenance
soft key to display the Maintenance menu.
4.
Press
the
Calibration Date
soft key. The following menu appears.
5.
Press the
Pressure Cal. Date
soft key. The following screen appears.
6.
Use the arrow (▲, ▼, ◄, and ►) keys to enter the date, and then press
ENTER
.
7.
When you calibrate the DMM function (/DM option) or D/A output (/DA option), press the DMM
Cal. Date or D/A Cal. Date soft key, and enter the calibration date, in the same manner as
Pressure Cal. Date.