49
IM 12D06D05-01E
A
APPENDICES
B. Calculation of temperature coefficient factor
(with two known conductivity values at different temperatures)
Measure the conductivity of the liquid at two temperatures, one below the reference and above
the reference temperature with the temperature coefficient set to 0,00%/°C and use the following
equation to calculate a temperature coefficient (α).
K
T
K
ref
=
1+
α
( T - T
ref
)
K
2
- K
1
α
= K
1
( T
2
- T
ref
) - K
2
( T
1
- T
ref
)
K
1
K
2
K
ref
=
=
1+
α
( T
1
- T
ref
)
1+
α
( T
2
- T
ref
)
K
1
(1 +
α
( T
2
- T
ref
)) = K
2
(1 +
α
( T
1
- T
ref
))
K
1
+
α
( T
2
- T
ref
) - K
2
+
α
( T
1
- T
ref
) = K
2
- K
1
Where T
1
, T
2
: liquid temperature (°C)
K
1
: conductivity at T
1
(°C)
K
2
: conductivity at T
2
(°C)
Figure 11-1. Conductivity
Calculation example
Calculate the temperature co-efficient of a liquid from the following data.
Conductivity 124.5 µS/cm at a liquid temperature of 18.0 °C and a
conductivity 147.6
µ
S/cm at a liquid temperature of 31.0 °C.
Substituting the data in the above formula gives the following result.
147.6 - 124.5
x 100= 1.298 %/C
α
=
124.5(31.0 - 25) - 147.6(18.0 - 25)
Set the temperature coefficient in the ISC450G converter.
Содержание ISC450G
Страница 1: ...IM 12D06D05 01E 4th Edition User s Manual Model ISC450G Style S2 Inductive Conductivity Converter...
Страница 8: ...vi...
Страница 23: ...15 IM 12D06D05 01E 3 INSTALLATION AND WIRING Figure 3 11 a Figure 3 11 b Figure 3 11 c...
Страница 54: ...46 IM 12D06D05 01E...
Страница 55: ...47 IM 12D06D05 01E 10 SPARE PARTS 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 66: ...58 IM 12D06D05 01E APPENDIX 6 Control drawing for FM approval...