YOKOGAWA ISC450G Скачать руководство пользователя страница 48

40 

IM 12D06D05-01E

7-1. Periodic maintenance 

The converter requires very little periodic main-

tenance, except to make sure the front window 

is kept clean in order to permit a clear view of 

the display and allow proper operation of the 

touchscreen. If the window becomes soiled, 

clean it using a soft damp cloth or soft tissue. 

To deal with more stubborn stains, a neutral 

detergent may be used.

When you must open the front cover and/or 

glands, make sure that the seals are clean and 

correctly fitted when the unit is re-assembled 

in order to maintain the housing’s weatherproof 

integrity against water and water vapor.

Note!

  Never use harsh chemicals or solvents. 

In the event that the window does 

become heavily stained or scratched, 

refer to the parts list (Chapter 10) for 

replacement part numbers.

Battery

The EXAxt converter contains a logbook 

feature that uses a clock to provide the timings. 

The instrument contains a lithium cell (battery) 

to support the clock function when the power 

is switched off. The cell has an expected work-

ing life of 10 years. Should this cell need to 

be replaced, contact your nearest Yokogawa 

service center.

Fuse

There is a circuit board mounted fuse protect-

ing the instrument. If you suspect that this 

needs to be replaced, contact your nearest 

Yokogawa service center.

7-2. Periodic maintenance of the sensor

Note!

  Maintenance advice listed here is 

intentionally general in nature. Sensor 

maintenance is highly application 

specific.

In general conductivity measurements do not 

need much periodic maintenance. If the EXAxt 

indicates an error in the measurement or in the 

calibration, some action may be needed (ref. 

chapter 8 troubleshooting).

7-3. Cleaning methods

1.

 For normal applications hot water with 

domestic washing-up liquid added will be 

effective.

2.

 For lime, hydroxides, etc., a 5 ...10% 

solution of hydrochloric acid is 

recommended.

3.

 Organic contaminants (oils, fats, etc.) can be 

easily removed with acetone.

4.

 For algae, bacteria or moulds, use a solution 

of domestic bleach (hypochlorite).

*  Never use hydrochloric acid and bleaching 

liquid simultaneously. The release of the very 

poisonous chlorine gas will result.

7-4. LCD adjustment

Contrast adjustment

During the life of the analyzer the contrast of 

the display may fade. The contrast can be ad-

justed using the potentiometer on the backside 

of the LCD board. This adjustment must be 

done only by Yokogawa’s service personnel. 

The position is shown on the picture below. For 

units manufactured after July 2007, the poten-

tiometer is placed behind the little hole in the 

LCD bracket as shown in Figure 3-4 on page 7.

For units manufactured between April 2006 

and April 2007, the potentiometer is located as 

shown below.

7. MAINTENANCE

Содержание ISC450G

Страница 1: ...IM 12D06D05 01E 4th Edition User s Manual Model ISC450G Style S2 Inductive Conductivity Converter...

Страница 2: ...be referred to when reading pages where subsequent submenu screens are shown in the text Connection to the relevant submenu screen is indicated by a doted line with an arrow Note that screens in the...

Страница 3: ...hat they are suitable the particular purpose of the user Every effort has been made to ensure accuracy in the preparation of this manual However when you realize mistaken expressions or omissions plea...

Страница 4: ...ganisations can deviate from the typical warranty period and the conditions of sale relating to the original purchase order should be consulted Dam age caused by wear and tear inadequate maintenance c...

Страница 5: ...conduit adapter 8 3 3 Wiring the power supply 9 3 3 1 General precautions 9 3 3 2 Access to terminal and cable entry 10 3 3 3 AC power 10 3 3 4 DC power 10 3 3 5 Grounding the housing 10 3 3 6 Switchi...

Страница 6: ...ftware revision 17 4 3 14 HART Device revision 18 4 3 15 Logbook 18 4 3 16 Trouble shooting 18 4 4 Information function 18 4 5 Setup Calibration commissioning 18 4 6 Secondary primary value display sw...

Страница 7: ...ng methods 40 7 4 LCD adjustment 40 8 TROUBLESHOOTING 42 8 1 General 42 8 2 Calibration check 42 8 3 Predictive maintenance 42 8 4 Poor calibration technique 42 8 5 Error displays and actions 42 9 QUA...

Страница 8: ...vi...

Страница 9: ...nductiv ity and or Concentration in industrial instal lations The unit combines simple operation and microprocessor based performance with advanced self diagnostics and enhanced communications capabil...

Страница 10: ...dded to above performance mA output tolerance 0 02 mA of 4 20 mA D Transmission signals General Two isolated outputs of 4 20 mA DC with common negative Maximum load 600 Bi directional HART digital com...

Страница 11: ...k Software record of important events and diagnostic data readily available in the display or through HART J Display Graphical Quarter VGA 320 x 240 pixels LCD with LED backlight and touchscreen Plain...

Страница 12: ...and operational conditions Ambient temperature 20 to 55 C 5 130 F Storage temperature 30 to 70 C 20 160 F Humidity 10 to 90 at 40 C 100 F RH non condensing Data protection EEPROM for configuration dat...

Страница 13: ...e limits of the instrument specifications See chapter 2 3 1 2 Mounting methods Refer to figures 3 2 and 3 3 Note that the EXAxt converter has universal mounting capa bilities Panel mounting using opti...

Страница 14: ...e 3 3 Option U wall and pipe mounting diagram 80 3 15 2x 6 5 0 26 4x 10 0 4 200 7 87 70 2 75 154 5 6 08 2 ND pipe pipe mounting horizontal pipe mounting vertical wall mounting 15 0 6 35 1 38 13 0 51 U...

Страница 15: ...terminals 6 Connect the analog output s the sensor inputs and if necessary the HART wiring and input contact 7 Use the front three glands for analog output sensor inputs contact input and HART wiring...

Страница 16: ...ands includ ing the blind plugs supplied ISC450G A D U The ISC450 is supplied with terminals suitable for the connection of finished wires in the size range of 14 26 AWG The cable entry holes are seal...

Страница 17: ...ower cables and contact outputs should be fixed securely onto a wall or construction using cable racks conduit tubing nylon bands or other appropriate ways Accidental removal from terminals by pulling...

Страница 18: ...OR S REFER TO INSTRUCTION MANUAL FOR CONNECTIONS ISC mA1 HART 14 COILS RECEIVE DRIVE SHLD 13 32 31 33 42 41 43 52 51 53 72 71 73 NC C NO NC C NO NC C NO NO C NC S1 S2 S3 CONTACTS S4 250V 5A AC DC 100V...

Страница 19: ...button on the main screen the user is given an explanation as well as a remedy for the current fail situation Always connect the fail contact to an alarm device such as a warning light alarm bell or...

Страница 20: ...r additional me chan i cal pro tec tion of the sensor cable 3 6 3 Installation of the sensor The Model ISC40 is a doughnut shaped sen sor Preferably the process will flow through the hole of the donut...

Страница 21: ...nufactured to a very high standard and are necessary to ensure that the specifications of the system can be met The total cable length should not exceed 60 metres e g 5 m fixed cable and 55 m exten si...

Страница 22: ...cores 5 Extract the coaxial cores from the braid and trim off the black low noise screening material as short as possible 6 Insulate the overall screen and the 2 coaxial screens with suitable plastic...

Страница 23: ...15 IM 12D06D05 01E 3 INSTALLATION AND WIRING Figure 3 11 a Figure 3 11 b Figure 3 11 c...

Страница 24: ...full screen displays a trend of 51 points that represent the average of the selected time interval The analyzer samples the measurement every second The trending graphic also shows the maximum and mi...

Страница 25: ...this reading and the C C factory value the sensor should be checked for damage and cleanliness Routing is via the Calibration menu 4 3 6 Temp comp 1 the chosen temperature compensation method for the...

Страница 26: ...he events logged in this logbook are user definable This is done in Commissioning Configure Logbook Settings Logbook mA and or Settings Logbook contact mA1 mA2 shows all dynamic events concern ing the...

Страница 27: ...19 IM 12D06D05 01E 4 OPERATION OF EXAxt ISC450G RETURN KEY exit to previous display Main display Primary setup display Commisioning menu display Instrument in HOLD 4 7 Navigation of the menu structure...

Страница 28: ...perature that represents the process tem perature must be set here Reference Temperature Choose a temperature to which the measured conductivity value must be compensated Normally 25 C 77 C is used th...

Страница 29: ...emp Comp Reference Temp 25 C 77 F 0 C 32 F 100 C 211 F Manual Comp Manual Temp 25 C 77 F 20 C 3 9 F 139 C 284 F Temp Coef T C methods 1 2 10 C 0 C 0 F 3 5 C 2 0 F Temp Coef T C methods 2 2 10 C 0 C 0...

Страница 30: ...bandwidth of 1 over a period of the stabilization time the calibration is considered stable and the calibration may be completed Calibration Interval A user defined interval in which a new calibra tio...

Страница 31: ...ization time 5 s 2 s 30 s Calib interval 250 days 1 day 250 days Concentr Table Table See appendix Enter Measurement setup Temperature settings Temp Compensation Calibration settings Concentration Mea...

Страница 32: ...desired setpoint The output signal is represented as a percentage of output 0 100 Proportional control will reduce but not eliminate the steady state error Therefore proportional Control action includ...

Страница 33: ...C F inf inf Linear mA2 100 Value 100 0 C F inf inf P control mA1 Setpoint 500 0 mS cm inf inf P control mA1 Range 100 0 mS cm inf inf P control mA1 Manual Reset 0 000 0 100 P control mA2 Setpoint 25...

Страница 34: ...wer on power down normal opened contact activated S1 S2 S3 S4 Above table shows contact output status between common to NO Configure hold Hold is the procedure to set the outputs to a known state when...

Страница 35: ...Alarm S2 Setpoint 100 0 mS cm low inf inf Alarm S2 Hysteresis 1 000 mS cm 0 0 S cm inf Alarm S2 Delay Time 0 2 sec 0 0 sec inf Alarm S2 Expire Time 0 0 sec 0 0 sec 1800 sec S1 control Expire Time 0 0...

Страница 36: ...rs 5 9 Simulate The contact can be switched on off or a percentage of output can be simulated On Off enables the user to manually switch a contact on or off The percentage is an analogue value and rep...

Страница 37: ...29 IM 12D06D05 01E 5 MENU STRUCTURE COMMISSIONING Menu Parameter Default Range values min max Simulation Percentage 50 0 100 S2 S3 S4 Similar structure to S1 depending on mA1 and mA2 output settings...

Страница 38: ...l and the converter continues to work normally A good example is a time out warning that the regular maintenance is due The user is noti fied but it should not be used to shut down the whole measureme...

Страница 39: ...31 IM 12D06D05 01E 5 MENU STRUCTURE COMMISSIONING Menu Parameter Default Range values min max Errors High limit 1 000S 0 1S 3S Errors Low limit 5 000 S 0 0S 100 S...

Страница 40: ...r Passwords Calibration and Commissioning may be sepa rately protected by a password By default both passwords are empty Entering an empty pass word results in disabling the password check A password...

Страница 41: ...33 IM 12D06D05 01E 5 MENU STRUCTURE COMMISSIONING Menu Parameter Default Range values Low High HART Network address 0 0 15...

Страница 42: ...d secondary display lines are determined by the choices made in the menu measurement Measurement setup Measurement Additional text Each process value can be given an additional text containing up to 1...

Страница 43: ...Parameter Default Range values Low High Y axis Conduct low 0 S cm inf inf Y axis Conduct high 500 S cm inf inf Y axis Conduct 2 low 0 S cm inf inf Y axis Conduct 2 high 500 S cm inf inf Y axis Temp lo...

Страница 44: ...ed with a known precise concentration Thetem per a ture is stabilized to the ref er ence tem per a ture of the instrument usu al ly 25 C The actual conductivity value of the solution is taken from tab...

Страница 45: ...ndicated on the cable markers and the actual installation can change this factor If there is less than 25 mm spacing between sensor and holder in situ calibration is necessary to meet the specified ac...

Страница 46: ...a better accuracy at low readings This should be done for all installations during commissioning After some time in service a dirty sensor may well show a high zero offset because of fouling Clean the...

Страница 47: ...ali bration Press Temperature calibration to execute it 6 8 Operation of hold function during cali bration EXAxt ISC450G has a HOLD function that will suspend the operation of the control alarm re lay...

Страница 48: ...ed contact your nearest Yokogawa service center 7 2 Periodic maintenance of the sensor Note Maintenance advice listed here is intentionally general in nature Sensor maintenance is highly application s...

Страница 49: ...ower then on again The touchscreen will be calibrated automatically to the correct touch position at power on It is recommended to turn off power then on again when periodic maintenance Do not turn on...

Страница 50: ...oting procedures including possible causes and remedies 8 2 Calibration check The EXAxt ISC450G converter incorporates a diagnostic check of the adjusted cell constant value during calibration If the...

Страница 51: ...rminal 3 Between the current output terminals shorted 62 and the protective earth terminal 3 2 Dielectric Strength Test 1 Apply 1390 V AC an AC voltage of substantially sinusoidal waveform with a freq...

Страница 52: ...wn in Table 2 Table 2 RANGE Resistance of Decade Resistance Box 2 Data Display RANGE4 1 5k 1 5k 0 01k RANGE5 15k 15k 0 2k RANGE6 150k 150k 12k After the above test is completed touch the Exit to retur...

Страница 53: ...ta Display Current Output mA DC 4 4 0 02 12 12 0 02 20 20 0 02 After all tests are completed a Touch the Exit twice to return to the Service display b Select Normal in Key c Touch the Home icon to ret...

Страница 54: ...46 IM 12D06D05 01E...

Страница 55: ...47 IM 12D06D05 01E 10 SPARE PARTS 10 SPARE PARTS See Customer Maintenance Parts List...

Страница 56: ...ard temperature compensation From the factory the EXAxt is set with the default of a general temperature compensation func tion based on a Sodium Chloride table salt solution This is suitable for many...

Страница 57: ...Tref K2 T1 Tref K1 K2 K ref 1 T1 Tref 1 T2 Tref K1 1 T2 Tref K2 1 T1 Tref K1 T2 Tref K2 T1 Tref K2 K1 Where T1 T2 liquid temperature C K1 conductivity at T1 C K2 conductivity at T2 C Figure 11 1 Condu...

Страница 58: ...r a particular solution Choose the component that you will be measuring in your application and where appro priate the concentration range EXAxt will do the rest By following the routing Commissioning...

Страница 59: ...ix value An empty value is shown as APPENDIX 2 Temperature compensation matrix 1 A minimum number of values is required to make interpolation possible The highlighted values marked as are mandatory to...

Страница 60: ...he calibration check can confirm their presence by a deviation from the known solution conductivity The remedial action should be to clean the sensor and carefully check for blockage or damage not sim...

Страница 61: ...l is of a high permeability mag net ic ma terial and a strong magnetic field is generated in the toroid The liquid passes through the hole in the toroid and can be considered as a one turn second ary...

Страница 62: ...om Zoom sensor Fact CC Adj CC Method SC1 Method SC2 Ohms Zoom outputs mA1 value mA2 value S1 perc S2 perc S3 perc S4 perc Zoom device Serial number Note Software Revision Device Revision DD Revision L...

Страница 63: ...C Calibration SC1 TC Calibration SC2 Temp Calibration Hold Instrument Hold Instrument Hold Outputs Hold Off Commissioning Measurement setup Configure sensor Meas unit Fact CC Measure Temp settings Tem...

Страница 64: ...o S1 S4 setup similar to S1 Type control Func Process parameter Expire time PID SP PID Rng PID dir PID MR PID I time PID D time Analog output DC period time max pulse freq Type fail func Type alarm Fu...

Страница 65: ...Device information Date Descriptor Message Poll addr Num resp preams Review Model Distributor Write protect Manufacturer Dev id Tag Descriptor Message Date Universal rev Fld dev rev Sofware rev Poll a...

Страница 66: ...58 IM 12D06D05 01E APPENDIX 6 Control drawing for FM approval...

Страница 67: ...5c K9676MY 1 Cable assembly shield 6a A1108EF 1 Fuse AC version 1 pcs 6b A1111EF 1 Fuse DC version 1 pcs 7 K9676BE 1 Cover assembly without screws and hingepins 8 K9676BU 1 Cable glands assembly 6 pcs...

Страница 68: ...ing Hardware Option code PM Item Part No Qty Description 1 K9171SS 1 Mounting Set U 2 Y9608KU 4 Screw 3 D0117XL A 1 U Bolt Assembly 4 K9171SY 1 Plate 5 K9171SX 1 Bracket 6 K9171ST 1 Mounting Set PM 7...

Страница 69: ...nged because external dimensions for awning hood H5 conduit adapter AFTG ANSI AM20 added p7 Figure 3 4 title modified p8 conduit adapter work added to subsection 3 2 2 p9 to 14 Layout changed descript...

Страница 70: ...Style S2 Inductive Conductivity Converter The user s manual IM 12D06D05 01E 4th Edition supplied with the product has been amended as follows Please replace the corresponding pages in your copy with t...

Страница 71: ...f the instrument specifications See chapter 2 3 1 2 Mounting methods Refer to figures 3 2 and 3 3 Note that the EXAxt converter has universal mounting capa bilities Panel mounting using optional brack...

Страница 72: ...3 Option U wall and pipe mounting diagram 2 ND pipe pipe mounting horizontal pipe mounting vertical wall mounting 15 0 6 35 1 38 13 0 51 4x 6 5 0 26 4x 10 0 4 200 7 87 70 2 75 154 5 6 08 80 3 15 80 3...

Страница 73: ...the Temperature elements used The other is a manual set temperature The manual temperature that represents the process tem perature must be set here Reference Temperature Choose a temperature to whic...

Отзывы: