5-6. mA output setup

The general procedure is to first define the 

function (control, output, simulate, off) of the 

output and second the process parameter 

associated to the output. Available process 

parameters depend on selected “sensor type” 

and “measurement setup”.

Off 

:  When an output is set off the  

output is not used and will give an 

output of 4 mA.

Control  :  A selection of P- PI- or PID control

Manual   : Static output required to maintain

reset 

  equilibrium state with setpoint.

Direction  :  Direct 

If the process variable is too high 

relative to the SP, the output of 

the controller is increased (direct 

action).

 

:   Reverse 

If the process variable is too high 

relative to the SP, the output of the 

controller is decreased (reverse 

action).

Output  :  Linear or non linear table output.

 

The table function allows the configu-

ration of an output curve by 21 steps 

(5% intervals). In the main menu 

concentration can be selected to set 

the concentration range.

Simulate  :   Percentage of output span.

 

Normal span of outputs are limited 

from 3.8 to 20.5 mA

Fail safe  :  Contact S4 is programmed as a 

 

fail-safe contact.

Burn Low or High will give an output of 3.6 

resp. 21 mA in case of Fail situation.

Note! 

When leaving Commissioning, Hold 

remains active until switched off 

manually. This is to avoid inappropriate 

actions while setting up the 

measurement.

Proportional control

Proportional Control action produces an output 

signal that is proportional to the difference 

between the Setpoint and the PV (deviation or 

error). Proportional control amplifies the error to 

motivate the process value towards the desired 

setpoint. The output signal is represented as a 

percentage of output (0-100%).

Proportional control will reduce but not 

eliminate the steady state error. Therefore, 

proportional Control action includes a Manual 

Reset. The manual reset (percentage of output) 

is used to eliminate the steady state error.

Note! 

Any changes (disturbances) in the 

process will re-introduce a steady state 

error.

Proportional control can also produce exces-

sive overshoot and oscillation. Too much gain 

may result in an unstable- or oscillating proc-

ess. Too little gain results in a sustained steady 

state error. 

Gain = 1/Range. [PV units]

Integral Control

Integral control is used to eliminate the steady 

state error and any future process changes. 

It will accumulate setpoint and process (load) 

changes by continuing to adjust the output 

until the error is eliminated. Small values of 

integral term (I-time in seconds) provide quick 

compensation, but increase overshoot. Usually, 

the integral term is set to a maximum value that 

provides a compromise between the three sys-

tem characteristics of: overshoot, settling time, 

and the time necessary to cancel the effects of 

static loading (process changes). The integral 

term is provided with an anti windup function. 

When the output of PI portion of the controller 

is outside the control range (less than -5% or 

greater than 105%), the I-part is frozen.

SP

PV

e

+-

++

++

+-

e

Range

ºe dt

1

T

i

T

d

dPV

dt

z

Process

Controller

Actuator

Process

Figure 5-1. Control Diagram

Derivative control

The control acts on the slope (rate of change) 

of the process value, thereby minimizing 

overshoot. It provides “rate” feedback, resulting 

in more damping. High derivative gains can 

increase the rizing time and settling time. It is 

difficult to realize in practice because differen-

tiation leads to “noisy” signals.

Содержание ISC450G

Страница 1: ...IM 12D06D05 01E 4th Edition User s Manual Model ISC450G Style S2 Inductive Conductivity Converter...

Страница 2: ...be referred to when reading pages where subsequent submenu screens are shown in the text Connection to the relevant submenu screen is indicated by a doted line with an arrow Note that screens in the...

Страница 3: ...hat they are suitable the particular purpose of the user Every effort has been made to ensure accuracy in the preparation of this manual However when you realize mistaken expressions or omissions plea...

Страница 4: ...ganisations can deviate from the typical warranty period and the conditions of sale relating to the original purchase order should be consulted Dam age caused by wear and tear inadequate maintenance c...

Страница 5: ...conduit adapter 8 3 3 Wiring the power supply 9 3 3 1 General precautions 9 3 3 2 Access to terminal and cable entry 10 3 3 3 AC power 10 3 3 4 DC power 10 3 3 5 Grounding the housing 10 3 3 6 Switchi...

Страница 6: ...ftware revision 17 4 3 14 HART Device revision 18 4 3 15 Logbook 18 4 3 16 Trouble shooting 18 4 4 Information function 18 4 5 Setup Calibration commissioning 18 4 6 Secondary primary value display sw...

Страница 7: ...ng methods 40 7 4 LCD adjustment 40 8 TROUBLESHOOTING 42 8 1 General 42 8 2 Calibration check 42 8 3 Predictive maintenance 42 8 4 Poor calibration technique 42 8 5 Error displays and actions 42 9 QUA...

Страница 8: ...vi...

Страница 9: ...nductiv ity and or Concentration in industrial instal lations The unit combines simple operation and microprocessor based performance with advanced self diagnostics and enhanced communications capabil...

Страница 10: ...dded to above performance mA output tolerance 0 02 mA of 4 20 mA D Transmission signals General Two isolated outputs of 4 20 mA DC with common negative Maximum load 600 Bi directional HART digital com...

Страница 11: ...k Software record of important events and diagnostic data readily available in the display or through HART J Display Graphical Quarter VGA 320 x 240 pixels LCD with LED backlight and touchscreen Plain...

Страница 12: ...and operational conditions Ambient temperature 20 to 55 C 5 130 F Storage temperature 30 to 70 C 20 160 F Humidity 10 to 90 at 40 C 100 F RH non condensing Data protection EEPROM for configuration dat...

Страница 13: ...e limits of the instrument specifications See chapter 2 3 1 2 Mounting methods Refer to figures 3 2 and 3 3 Note that the EXAxt converter has universal mounting capa bilities Panel mounting using opti...

Страница 14: ...e 3 3 Option U wall and pipe mounting diagram 80 3 15 2x 6 5 0 26 4x 10 0 4 200 7 87 70 2 75 154 5 6 08 2 ND pipe pipe mounting horizontal pipe mounting vertical wall mounting 15 0 6 35 1 38 13 0 51 U...

Страница 15: ...terminals 6 Connect the analog output s the sensor inputs and if necessary the HART wiring and input contact 7 Use the front three glands for analog output sensor inputs contact input and HART wiring...

Страница 16: ...ands includ ing the blind plugs supplied ISC450G A D U The ISC450 is supplied with terminals suitable for the connection of finished wires in the size range of 14 26 AWG The cable entry holes are seal...

Страница 17: ...ower cables and contact outputs should be fixed securely onto a wall or construction using cable racks conduit tubing nylon bands or other appropriate ways Accidental removal from terminals by pulling...

Страница 18: ...OR S REFER TO INSTRUCTION MANUAL FOR CONNECTIONS ISC mA1 HART 14 COILS RECEIVE DRIVE SHLD 13 32 31 33 42 41 43 52 51 53 72 71 73 NC C NO NC C NO NC C NO NO C NC S1 S2 S3 CONTACTS S4 250V 5A AC DC 100V...

Страница 19: ...button on the main screen the user is given an explanation as well as a remedy for the current fail situation Always connect the fail contact to an alarm device such as a warning light alarm bell or...

Страница 20: ...r additional me chan i cal pro tec tion of the sensor cable 3 6 3 Installation of the sensor The Model ISC40 is a doughnut shaped sen sor Preferably the process will flow through the hole of the donut...

Страница 21: ...nufactured to a very high standard and are necessary to ensure that the specifications of the system can be met The total cable length should not exceed 60 metres e g 5 m fixed cable and 55 m exten si...

Страница 22: ...cores 5 Extract the coaxial cores from the braid and trim off the black low noise screening material as short as possible 6 Insulate the overall screen and the 2 coaxial screens with suitable plastic...

Страница 23: ...15 IM 12D06D05 01E 3 INSTALLATION AND WIRING Figure 3 11 a Figure 3 11 b Figure 3 11 c...

Страница 24: ...full screen displays a trend of 51 points that represent the average of the selected time interval The analyzer samples the measurement every second The trending graphic also shows the maximum and mi...

Страница 25: ...this reading and the C C factory value the sensor should be checked for damage and cleanliness Routing is via the Calibration menu 4 3 6 Temp comp 1 the chosen temperature compensation method for the...

Страница 26: ...he events logged in this logbook are user definable This is done in Commissioning Configure Logbook Settings Logbook mA and or Settings Logbook contact mA1 mA2 shows all dynamic events concern ing the...

Страница 27: ...19 IM 12D06D05 01E 4 OPERATION OF EXAxt ISC450G RETURN KEY exit to previous display Main display Primary setup display Commisioning menu display Instrument in HOLD 4 7 Navigation of the menu structure...

Страница 28: ...perature that represents the process tem perature must be set here Reference Temperature Choose a temperature to which the measured conductivity value must be compensated Normally 25 C 77 C is used th...

Страница 29: ...emp Comp Reference Temp 25 C 77 F 0 C 32 F 100 C 211 F Manual Comp Manual Temp 25 C 77 F 20 C 3 9 F 139 C 284 F Temp Coef T C methods 1 2 10 C 0 C 0 F 3 5 C 2 0 F Temp Coef T C methods 2 2 10 C 0 C 0...

Страница 30: ...bandwidth of 1 over a period of the stabilization time the calibration is considered stable and the calibration may be completed Calibration Interval A user defined interval in which a new calibra tio...

Страница 31: ...ization time 5 s 2 s 30 s Calib interval 250 days 1 day 250 days Concentr Table Table See appendix Enter Measurement setup Temperature settings Temp Compensation Calibration settings Concentration Mea...

Страница 32: ...desired setpoint The output signal is represented as a percentage of output 0 100 Proportional control will reduce but not eliminate the steady state error Therefore proportional Control action includ...

Страница 33: ...C F inf inf Linear mA2 100 Value 100 0 C F inf inf P control mA1 Setpoint 500 0 mS cm inf inf P control mA1 Range 100 0 mS cm inf inf P control mA1 Manual Reset 0 000 0 100 P control mA2 Setpoint 25...

Страница 34: ...wer on power down normal opened contact activated S1 S2 S3 S4 Above table shows contact output status between common to NO Configure hold Hold is the procedure to set the outputs to a known state when...

Страница 35: ...Alarm S2 Setpoint 100 0 mS cm low inf inf Alarm S2 Hysteresis 1 000 mS cm 0 0 S cm inf Alarm S2 Delay Time 0 2 sec 0 0 sec inf Alarm S2 Expire Time 0 0 sec 0 0 sec 1800 sec S1 control Expire Time 0 0...

Страница 36: ...rs 5 9 Simulate The contact can be switched on off or a percentage of output can be simulated On Off enables the user to manually switch a contact on or off The percentage is an analogue value and rep...

Страница 37: ...29 IM 12D06D05 01E 5 MENU STRUCTURE COMMISSIONING Menu Parameter Default Range values min max Simulation Percentage 50 0 100 S2 S3 S4 Similar structure to S1 depending on mA1 and mA2 output settings...

Страница 38: ...l and the converter continues to work normally A good example is a time out warning that the regular maintenance is due The user is noti fied but it should not be used to shut down the whole measureme...

Страница 39: ...31 IM 12D06D05 01E 5 MENU STRUCTURE COMMISSIONING Menu Parameter Default Range values min max Errors High limit 1 000S 0 1S 3S Errors Low limit 5 000 S 0 0S 100 S...

Страница 40: ...r Passwords Calibration and Commissioning may be sepa rately protected by a password By default both passwords are empty Entering an empty pass word results in disabling the password check A password...

Страница 41: ...33 IM 12D06D05 01E 5 MENU STRUCTURE COMMISSIONING Menu Parameter Default Range values Low High HART Network address 0 0 15...

Страница 42: ...d secondary display lines are determined by the choices made in the menu measurement Measurement setup Measurement Additional text Each process value can be given an additional text containing up to 1...

Страница 43: ...Parameter Default Range values Low High Y axis Conduct low 0 S cm inf inf Y axis Conduct high 500 S cm inf inf Y axis Conduct 2 low 0 S cm inf inf Y axis Conduct 2 high 500 S cm inf inf Y axis Temp lo...

Страница 44: ...ed with a known precise concentration Thetem per a ture is stabilized to the ref er ence tem per a ture of the instrument usu al ly 25 C The actual conductivity value of the solution is taken from tab...

Страница 45: ...ndicated on the cable markers and the actual installation can change this factor If there is less than 25 mm spacing between sensor and holder in situ calibration is necessary to meet the specified ac...

Страница 46: ...a better accuracy at low readings This should be done for all installations during commissioning After some time in service a dirty sensor may well show a high zero offset because of fouling Clean the...

Страница 47: ...ali bration Press Temperature calibration to execute it 6 8 Operation of hold function during cali bration EXAxt ISC450G has a HOLD function that will suspend the operation of the control alarm re lay...

Страница 48: ...ed contact your nearest Yokogawa service center 7 2 Periodic maintenance of the sensor Note Maintenance advice listed here is intentionally general in nature Sensor maintenance is highly application s...

Страница 49: ...ower then on again The touchscreen will be calibrated automatically to the correct touch position at power on It is recommended to turn off power then on again when periodic maintenance Do not turn on...

Страница 50: ...oting procedures including possible causes and remedies 8 2 Calibration check The EXAxt ISC450G converter incorporates a diagnostic check of the adjusted cell constant value during calibration If the...

Страница 51: ...rminal 3 Between the current output terminals shorted 62 and the protective earth terminal 3 2 Dielectric Strength Test 1 Apply 1390 V AC an AC voltage of substantially sinusoidal waveform with a freq...

Страница 52: ...wn in Table 2 Table 2 RANGE Resistance of Decade Resistance Box 2 Data Display RANGE4 1 5k 1 5k 0 01k RANGE5 15k 15k 0 2k RANGE6 150k 150k 12k After the above test is completed touch the Exit to retur...

Страница 53: ...ta Display Current Output mA DC 4 4 0 02 12 12 0 02 20 20 0 02 After all tests are completed a Touch the Exit twice to return to the Service display b Select Normal in Key c Touch the Home icon to ret...

Страница 54: ...46 IM 12D06D05 01E...

Страница 55: ...47 IM 12D06D05 01E 10 SPARE PARTS 10 SPARE PARTS See Customer Maintenance Parts List...

Страница 56: ...ard temperature compensation From the factory the EXAxt is set with the default of a general temperature compensation func tion based on a Sodium Chloride table salt solution This is suitable for many...

Страница 57: ...Tref K2 T1 Tref K1 K2 K ref 1 T1 Tref 1 T2 Tref K1 1 T2 Tref K2 1 T1 Tref K1 T2 Tref K2 T1 Tref K2 K1 Where T1 T2 liquid temperature C K1 conductivity at T1 C K2 conductivity at T2 C Figure 11 1 Condu...

Страница 58: ...r a particular solution Choose the component that you will be measuring in your application and where appro priate the concentration range EXAxt will do the rest By following the routing Commissioning...

Страница 59: ...ix value An empty value is shown as APPENDIX 2 Temperature compensation matrix 1 A minimum number of values is required to make interpolation possible The highlighted values marked as are mandatory to...

Страница 60: ...he calibration check can confirm their presence by a deviation from the known solution conductivity The remedial action should be to clean the sensor and carefully check for blockage or damage not sim...

Страница 61: ...l is of a high permeability mag net ic ma terial and a strong magnetic field is generated in the toroid The liquid passes through the hole in the toroid and can be considered as a one turn second ary...

Страница 62: ...om Zoom sensor Fact CC Adj CC Method SC1 Method SC2 Ohms Zoom outputs mA1 value mA2 value S1 perc S2 perc S3 perc S4 perc Zoom device Serial number Note Software Revision Device Revision DD Revision L...

Страница 63: ...C Calibration SC1 TC Calibration SC2 Temp Calibration Hold Instrument Hold Instrument Hold Outputs Hold Off Commissioning Measurement setup Configure sensor Meas unit Fact CC Measure Temp settings Tem...

Страница 64: ...o S1 S4 setup similar to S1 Type control Func Process parameter Expire time PID SP PID Rng PID dir PID MR PID I time PID D time Analog output DC period time max pulse freq Type fail func Type alarm Fu...

Страница 65: ...Device information Date Descriptor Message Poll addr Num resp preams Review Model Distributor Write protect Manufacturer Dev id Tag Descriptor Message Date Universal rev Fld dev rev Sofware rev Poll a...

Страница 66: ...58 IM 12D06D05 01E APPENDIX 6 Control drawing for FM approval...

Страница 67: ...5c K9676MY 1 Cable assembly shield 6a A1108EF 1 Fuse AC version 1 pcs 6b A1111EF 1 Fuse DC version 1 pcs 7 K9676BE 1 Cover assembly without screws and hingepins 8 K9676BU 1 Cable glands assembly 6 pcs...

Страница 68: ...ing Hardware Option code PM Item Part No Qty Description 1 K9171SS 1 Mounting Set U 2 Y9608KU 4 Screw 3 D0117XL A 1 U Bolt Assembly 4 K9171SY 1 Plate 5 K9171SX 1 Bracket 6 K9171ST 1 Mounting Set PM 7...

Страница 69: ...nged because external dimensions for awning hood H5 conduit adapter AFTG ANSI AM20 added p7 Figure 3 4 title modified p8 conduit adapter work added to subsection 3 2 2 p9 to 14 Layout changed descript...

Страница 70: ...Style S2 Inductive Conductivity Converter The user s manual IM 12D06D05 01E 4th Edition supplied with the product has been amended as follows Please replace the corresponding pages in your copy with t...

Страница 71: ...f the instrument specifications See chapter 2 3 1 2 Mounting methods Refer to figures 3 2 and 3 3 Note that the EXAxt converter has universal mounting capa bilities Panel mounting using optional brack...

Страница 72: ...3 Option U wall and pipe mounting diagram 2 ND pipe pipe mounting horizontal pipe mounting vertical wall mounting 15 0 6 35 1 38 13 0 51 4x 6 5 0 26 4x 10 0 4 200 7 87 70 2 75 154 5 6 08 80 3 15 80 3...

Страница 73: ...the Temperature elements used The other is a manual set temperature The manual temperature that represents the process tem perature must be set here Reference Temperature Choose a temperature to whic...

Отзывы: