<4. DETAILS OF MESSAGE>
4-2
IM 11G02Q02-01EN 2nd Edition : Sep. 04, 2020-00
(3) Message transmission (example)
The following shows an example of reading out from No. 1 station the setting Ch2 (2nd
component) range-1 zero and span calibration concentration.
Relative address of Ch2 range-1 zero calibration concentration setting: 0004
H
Data number: 02
H
Command message
Response message
Station No.
01
H
Station No.
01
H
Function code
03
H
Function code
03
H
Read-out start No.
(relative address)
Upper
00
H
Read-out byte number
04
H
Lower
04
H
Contents of the first
word data
Upper
00
H
Read-out word
number
Upper
00
H
Lower
00
H
Lower
02
H
Contents of the
next word data
Upper
03
H
CRC data
Lower
85
H
Lower
E8
H
Upper
CA
H
CRC data
Lower
FA
H
Upper
8D
H
•
The following describes how read-out data is compiled.
Ch2 range-1 zero calibration concentration setting
00. 00H = 0
(contents of first word data)
Ch2 range-1 span calibration concentration setting
03 E8H = 1000
(contents of next word data)
Provided decimal point position = 1, measurement unit = 1,
Ch2 range-1 zero calibration concentration setting =
0.0 ppm
Ch2 range-1 span calibration concentration setting =
100.0 ppmI
For handling of decimal point and unit, refer to “5.1.1 Handling of decimal point position
4.2
Read-out word data (Read-out only)
(Function code : 04
H
)
Function
code
Max.word
number read-out
in one message
Relative data
address
Register No.
Description
04
H
64 words
0000
H
―00C1
H
30001―30194
Measurement value and status
0425
H
―0469
H
31062―31130
Fixed setting