9.6 Error Log Display
Procedure
Error Log Display
1.
Press
UTILITY
to display the following menu.
2.
Press the
Error Log
soft key to display the error log.
Error description
Displayed when the error log
contains six lines or more.
3.
If the error log contains six lines or more, press the soft key to scroll the display.
Clearing the Error Log Display
4.
Press the
Clear
soft key to clear the error log.
Explanation
The ERROR indicator illuminates if an error such as a runtime error or communication
command error is stored in the error memory.
The error log displays errors and messages. An error has an associated error number,
and a message does not.
A message is displayed in the error log until the problem indicated by the message is
resolved. It is not cleared even if the Clear soft key is pressed.
For the meaning of errors and their corrective actions, see section 14.2, “Error Code
Descriptions and Corrective Actions.”
Note
• The error memory is also cleared if you use a communication command (“
:SYSTem:
ERRor?
”) to read errors, or if the GS200 is turned off. You can also clear the error memory
by sending the
*CLS
communication command.
• If the error memory becomes full, “
–350: Queue overflow
” is entered as the last error
indicating that any additional errors will be discarded. When the error memory is cleared, the
ERROR indicator turns off.
<<Corresponding Command Mnemonic>>
:SYSTem:ERRor?
:STATus:ERRor?
*CLS
9-7