background image

IM 12B06J03-05E-E

8

2. GENERAL SPECIFICATIONS

2.1 Measuring elements

 

 

: Na-glass electrode

 

 

: pH-glass electrode

 

 

: Silver chloride reference

 

 

: Solid platinum electrode

 

 

: Pt1000 temperature sensor. 

2.2 Wetted parts

Sensor body 

: PVDF-(GF25+TZ4)

Earthing pin 

: Solid Platinum

Measuring sensor 

: L-glass, pNa-glass

LE glass tube 

: AR-glass 

Sealing 

: Viton -FTS 

 

FFKM, EPDM  -MTS

Body insert 

: PVDF

2.3 Functional specifications (at 25°C)

Isothermal point 

: pH7, pNa 0

Reference system 

: Salt sensitive, Ag/AgCl in 1M KCl

Glass impedances 

: Nominal: 750 M

Ω

Liquid outlet 

: Non-flow no junction

Temperature element 

: Pt1000 to IEC 751

Asymmetry potential 

: 0 ± 15 mV

Linearity PH (Slope) 

: > 90% in pH 2-12 with pH = pNa+2

NOTE:

  The temperature sensor included in the FU20-FTS/MTS is designed for cell 

compensation and for indication. It is 

NOT

 designed for process temperature 

control.

2.4 Dynamic specifications

Response time pH  

: t90 < 15 sec. (for 7 to 4 pH step)

Response time temp. 

: t90 < 120 sec. (for 10 °C step)

Stabilization time pH 

: < 2 min. (for 0.02 pH unit during 10 sec.)

2.5 Operating range

pH   

: 2 to 14

ORP 

: -1500 to 1500 mV

Temperature 

: 0 to 105 ºC (14 to 221 ºF)

Pressure 

:   p(bar)

 
Conductivity 

: > 10 µS/cm

NOTE:

  The pH operating range at room temperature is 2-14 pH, but at high temperatures 

or range outside 2-12 pH the lifetime will be seriously shortened.

p(bar)

Temp. (ºC)

-10

25

50

105

0

5

0

10

FSM

Содержание FU20-MTS

Страница 1: ...IM 12B06J03 05E E 4th edition Instruction Manual Model FU20 FTS MTS Differential pH ORP sensor...

Страница 2: ...IM 12B06J03 05E E 2...

Страница 3: ...IM 12B06J03 05E E 3...

Страница 4: ...IM 12B06J03 05E E 4...

Страница 5: ...ON OF FU20 10 3 1 Typical installation 10 3 1 1 Preparing the sensor for use 10 3 1 2 Mounting the sensor 10 3 2 Mounting FU20 FTS using quick removal adapters 11 3 3 Mounting the FU20 FTS in PR10 ret...

Страница 6: ...ode and Serial Number on the sensor are the same as on the packing list Also check any option s that were ordered are included and correct 1 3 Warranty and Service Yokogawa products and parts are guar...

Страница 7: ...or year month numbering Table 1 Production Year code Table 2 Production Month code Year Year code Year Year code Month Month code 2014 P 2026 3 January 1 2015 R 2027 4 February 2 2016 S 2028 5 March 3...

Страница 8: ...e element Pt1000 to IEC 751 Asymmetry potential 0 15 mV Linearity PH Slope 90 in pH 2 12 with pH pNa 2 NOTE The temperature sensor included in the FU20 FTS MTS is designed for cell compensation and fo...

Страница 9: ...and control instruments ion selective electrodes Pressure Directive 97 23 EC as amended by Regulation EC no 1882 2003 Applying article 3 3 Sound Engineering Practice Damaging the screw thread of the s...

Страница 10: ...plane to eliminate air bubbles in the pH glass bulb see Figure 1 Figure 1 Sensor installation 3 1 1 Preparing the sensor for use Remove the sensor from its shipping box and slide of the so called wet...

Страница 11: ...2 Install the adapter in the process connection Tighten the adapter using a wrench on the adapter flats 3 Apply Teflon tape to the appropriate threaded end of the sensor 4 Place the O ring and screw t...

Страница 12: ...t undo the part in the metal tube 5 Lead the sensor cable through the tube of the fitting from the side where the knurled knob has been removed Attach the sensor and cable as usual 6 Hold the sensor s...

Страница 13: ...2B06J03 05E E 13 FU20 O ring Fitting Adapter FPS O ring HCNF SPRAY CLEANER FU20 FTS MTS ADAPTER FU20 FTS MTS FD40 Figure 5 Installation examples for the FU20 Process Flow Cleaning Solution water acid...

Страница 14: ...0 2 5 0 0 2 25 0 1 Figure 7 Dimensions 1 FU20 FTS MTS adapter Stainless Steel Titanium and FU20 FTS MTS adapter for FF40 FS40 and FD40 fittings 4 DIMENSIONS Figure 6 Dimensions FU20 FTS MTS FU20 VP T...

Страница 15: ...Yokogawa pH instrument Differential pH pNA ORP rH 15 16 13 17 11 12 14 63 3 none Differential ORP for pH 13 17 11 12 14 15 63 3 none Differential ORP 13 17 11 12 14 15 63 3 none pH compensated for pH...

Страница 16: ...ransparent Coax 2 Transparent Coax 2 Clear 14 Yellow 15 Transparent Coax 1 Yellow Yellow 16 Brown 17White Transparent Coax 2 Transparent Coax 2 Brown Note For ORP measurement the EXA FLXA terminals 13...

Страница 17: ...commended Set the pH analyzer to MANUAL calibration Wait until the pH reading is stable Adjust the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2...

Страница 18: ...FU20 FTS MTS sensor does not have an open reference junction and is hermetically sealed from the process it does not suffer from poisoning diffusion and fouling This means the FU20 FTS MTS sensor requ...

Страница 19: ...del NPT PPS body Tapered Thread Dome shaped FSM PPS body Tapered Thread Flat Surface FTD PVDF body Tapered Thread Dome shaped FTS PVDF body Tapered Thread Salt Sensitive membrane Silicone sealing MTS...

Страница 20: ...F10 xxx F Extension cable with terminated ends Specify length in whole meters e g XXX 005 010 025 050 meter WU10 V D 02 Cable for Variopin dual coax 2 meter WU10 V D 05 Cable for Variopin dual coax 5...

Страница 21: ...I chloride 50 O O O O O O O O O O O O O O O O O O Sodium sulfite sat O O O O O O O O O O O O O O O O O O Sodium carbonate sat O O O O O O O O O O O O O O O O O O O O O Potassium chloride sat O O O X X...

Страница 22: ...IM 12B06J03 05E E 22...

Страница 23: ...IM 12B06J03 05E E 23...

Страница 24: ...etherlands www yokogawa com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road New...

Отзывы: