IM 12B06J03-05E-E
11
3.2 Mounting FU20-FTS/MTS using
quick removal adapters
The FU20-FTS/MTS can also be mounted
using
one of the optional quick-removal adapters
(/NSS, /NTI, /BSS, /BTI, see Figure 3). For
a detailed description of these adapters see
Sections 4 and 7 of this Instruction Manual.
1 Apply Teflon tape to the threaded end of
this adapter;
2 Install the adapter in the process
connection. Tighten the adapter using a
wrench on the adapter flats.
3 Apply Teflon tape to the appropriate
threaded end of the sensor;
4 Place the O-ring and screw the mating
part of the adapter on the sensor;
5 Mount the sensor in the adapter, making
sure that the O-ring seals properly;
6 Hand-tighten the adapter nut.
Other mounting examples of the FU20-FTS
are given in Figure 4.
Note:
DO NOT
over tighten the adapter to
prevent damage;
Figure 3: Mounting of sensor with option /NSS, /NTI, /BSS or /BTI
1
2
3
4
5
6