background image

IM 12B06J03-05E-E

18

6.3  Calibration of ORP and rH 

measurements

For calibration of ORP and rH, the 
procedure for MANUAL CALIBRATION can 
be used as described in the Instruction 
Manual of the analyzer.

The rH value is a function of the reference 
system and the pH value of the buffer 
solution. The FU20-FTS/MTS sensor has 
a reference system of 1molal Silver/Silver 
Chloride (Ag/AgCl). The commonly used 
standards for ORP and rH calibration are 
made from Chinhydron (Quinhydrone) 
powder dissolved in pH buffer solutions (1 
g / 200 ml). In Table 5 the measurement 
values are given as function of the used 
pH buffer solution with Chinhydron 
powder. The accuracy of the standards is 
approximately ± 10 mV.

Table 5

: ORP, pH compensated ORP and 

rH as function of pH buffer solution with 
Chinhydron powder.

pH 

ORP (mV) Na  ORP (mV) pH  rH (mV)

buffer  compensated compensated

4.00 228 

51 

23.6

7.00 

51 51 23.6

9.00 -67 

51 

23.6

The accuracy of the standard values is 
approxi/- 10 mV.

6.4 Maintenance of the FU20-FTS/MTS 
sensor

A pH sensor requires routine maintenance 
to keep the measuring elements clean and 
functioning. Because the FU20-FTS/MTS 
sensor does not have an open reference 
junction and is hermetically sealed from the 
process, it does not suffer from poisoning, 
diffusion and fouling. This means the 
FU20-FTS/MTS sensor requires very little 
maintenance.

In most cases cleaning with water, iso-
propanol or methanol is sufficient. In other 
cases the measuring elements of the sensor 
have to be cleaned with specific solutions. 
Examples:
1. Deposits of limes, hydroxides or 

carbonates can be removed by 
immersing the bottom part of the sensor 
in a solution containing dilute hydrochloric 
acid (5% is recommended). Afterwards 
rinse the sensor with water.

2. Deposits of oil and fat can be removed 

with hot water with a detergent. When 
the results are unsatisfactory, a mild 
(carbonate based) abrasive can be used.

3. Protein deposits should be removed with 

a protein enzymatic solution, for instance 
a solution containing 8.5 ml concentrated 
hydrochloric acid and 10 grams of pepsin 
in 1 liter of water.

A soft toothbrush may be used to 

accelerate the cleaning process.

Note:

  Avoid cleaning with non-polar solvent 

like tri-chloro ethylene, toluene or 
hexane. The non-polar solvents 
will break up the gel-layer on the 
measuring glass elements and 
requires that the sensor has to be 
soaked in water for at least  
12 hours before it will function again.

Содержание FU20-MTS

Страница 1: ...IM 12B06J03 05E E 4th edition Instruction Manual Model FU20 FTS MTS Differential pH ORP sensor...

Страница 2: ...IM 12B06J03 05E E 2...

Страница 3: ...IM 12B06J03 05E E 3...

Страница 4: ...IM 12B06J03 05E E 4...

Страница 5: ...ON OF FU20 10 3 1 Typical installation 10 3 1 1 Preparing the sensor for use 10 3 1 2 Mounting the sensor 10 3 2 Mounting FU20 FTS using quick removal adapters 11 3 3 Mounting the FU20 FTS in PR10 ret...

Страница 6: ...ode and Serial Number on the sensor are the same as on the packing list Also check any option s that were ordered are included and correct 1 3 Warranty and Service Yokogawa products and parts are guar...

Страница 7: ...or year month numbering Table 1 Production Year code Table 2 Production Month code Year Year code Year Year code Month Month code 2014 P 2026 3 January 1 2015 R 2027 4 February 2 2016 S 2028 5 March 3...

Страница 8: ...e element Pt1000 to IEC 751 Asymmetry potential 0 15 mV Linearity PH Slope 90 in pH 2 12 with pH pNa 2 NOTE The temperature sensor included in the FU20 FTS MTS is designed for cell compensation and fo...

Страница 9: ...and control instruments ion selective electrodes Pressure Directive 97 23 EC as amended by Regulation EC no 1882 2003 Applying article 3 3 Sound Engineering Practice Damaging the screw thread of the s...

Страница 10: ...plane to eliminate air bubbles in the pH glass bulb see Figure 1 Figure 1 Sensor installation 3 1 1 Preparing the sensor for use Remove the sensor from its shipping box and slide of the so called wet...

Страница 11: ...2 Install the adapter in the process connection Tighten the adapter using a wrench on the adapter flats 3 Apply Teflon tape to the appropriate threaded end of the sensor 4 Place the O ring and screw t...

Страница 12: ...t undo the part in the metal tube 5 Lead the sensor cable through the tube of the fitting from the side where the knurled knob has been removed Attach the sensor and cable as usual 6 Hold the sensor s...

Страница 13: ...2B06J03 05E E 13 FU20 O ring Fitting Adapter FPS O ring HCNF SPRAY CLEANER FU20 FTS MTS ADAPTER FU20 FTS MTS FD40 Figure 5 Installation examples for the FU20 Process Flow Cleaning Solution water acid...

Страница 14: ...0 2 5 0 0 2 25 0 1 Figure 7 Dimensions 1 FU20 FTS MTS adapter Stainless Steel Titanium and FU20 FTS MTS adapter for FF40 FS40 and FD40 fittings 4 DIMENSIONS Figure 6 Dimensions FU20 FTS MTS FU20 VP T...

Страница 15: ...Yokogawa pH instrument Differential pH pNA ORP rH 15 16 13 17 11 12 14 63 3 none Differential ORP for pH 13 17 11 12 14 15 63 3 none Differential ORP 13 17 11 12 14 15 63 3 none pH compensated for pH...

Страница 16: ...ransparent Coax 2 Transparent Coax 2 Clear 14 Yellow 15 Transparent Coax 1 Yellow Yellow 16 Brown 17White Transparent Coax 2 Transparent Coax 2 Brown Note For ORP measurement the EXA FLXA terminals 13...

Страница 17: ...commended Set the pH analyzer to MANUAL calibration Wait until the pH reading is stable Adjust the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2...

Страница 18: ...FU20 FTS MTS sensor does not have an open reference junction and is hermetically sealed from the process it does not suffer from poisoning diffusion and fouling This means the FU20 FTS MTS sensor requ...

Страница 19: ...del NPT PPS body Tapered Thread Dome shaped FSM PPS body Tapered Thread Flat Surface FTD PVDF body Tapered Thread Dome shaped FTS PVDF body Tapered Thread Salt Sensitive membrane Silicone sealing MTS...

Страница 20: ...F10 xxx F Extension cable with terminated ends Specify length in whole meters e g XXX 005 010 025 050 meter WU10 V D 02 Cable for Variopin dual coax 2 meter WU10 V D 05 Cable for Variopin dual coax 5...

Страница 21: ...I chloride 50 O O O O O O O O O O O O O O O O O O Sodium sulfite sat O O O O O O O O O O O O O O O O O O Sodium carbonate sat O O O O O O O O O O O O O O O O O O O O O Potassium chloride sat O O O X X...

Страница 22: ...IM 12B06J03 05E E 22...

Страница 23: ...IM 12B06J03 05E E 23...

Страница 24: ...etherlands www yokogawa com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road New...

Отзывы: