6.3 Calibration of ORP and rH measurements 

For calibration of ORP and rH, the procedure for MANUAL CALIBRATION can be used as 
described in the Instruction Manual of the analyzer. The rH value is a function of the   
reference system and the pH value of the buffer solution. The FU20-FTS/MTS sensor has 
a reference system of 1molal Silver/Silver Chloride (Ag/AgCl). The commonly used 
standards for ORP and rH calibration are made from Chinhydron (Quinhydrone) powder 
dissolved in pH buffer solutions (1g / 200 ml). In Table 5 the measurement values are 
given as function of the used pH buffer solution with Chinhydron powder. The accuracy of 
the standards is approximately ± 10 mV. 
 

Table 8: ORP, pH compensated ORP and rH as function of pH buffer solution with 
Chinhydron powder. 

 

 
 

6.4 Maintenance of the FU20-FTS/MTS sensor 

A pH sensor requires routine maintenance to keep the measuring elements clean and 
functioning. Because the FU20-FTS/MTS sensor does not have an open reference 
junction and is hermetically sealed from the process, it does not suffer from poisoning, 
diffusion and fouling. This means the FU20-FTS/MTS sensor requires very little 
maintenance. 

 

Avoid cleaning the complete sensor with solution. Some cleaning solutions will 
damage the model code sticker and connector which are placed on top of the 
sensor. Only clean the measuring elements at the bottom side of the sensor.  

 
In most cases cleaning with water, iso-propanol or methanol is sufficient. In other cases, 
the measuring elements of the sensor have to be cleaned with specific solutions. 
 
Examples: 
1. Deposits of limes, hydroxides or carbonates can be removed by immersing the bottom 

part of the sensor in a solution containing dilute hydrochloric acid (5% is recommended). 
Afterwards rinse the sensor with water.  

2. Deposits of oil and fat can be removed with hot water with a detergent. When the results 

are unsatisfactory, a mild (carbonate based) abrasive can be used.  

3. Protein deposits should be removed with a protein enzymatic solution, for instance a 

solution containing 8.5 mL concentrated hydrochloric acid and 10 grams of pepsin in 1 
liter of water.  

A soft toothbrush may be used to accelerate the cleaning process. 
 

Note:  

Avoid cleaning with non-polar solvent like tri-chloro ethylene, toluene or hexane. 
The non-polar solvents will break up the gel-layer on the pH glass bulb and 
requires that the sensor has to be soaked in water for at least 12 hours before it 
will function again.  

 

 

 

pH buffer 

ORP (mV) Na 

ORP (mV) pH 

compensated 

rH (mV) 

4.00 

228 

51 

23.6 

7.00 

51 

51 

23.6 

9.00 

-67 

51 

23.6 

Содержание FU20-FTS

Страница 1: ...IM 12B06J03 05EN P 1st edition Instruction Manual Model FU20 FTS MTS Differential pH ORP sensor...

Страница 2: ...oole E Todos los manuales de instrucciones para los productos antiexplosivos de ATEX est n disponibles en ingl s Si desea solicitar las instrucciones de estos art culos antiexplosivos en su idioma loc...

Страница 3: ...com a delega o mais pr xima ou com um representante da Yokogawa PL Wszystkie instrukcje obs ugi dla urz dze w wykonaniu przeciwwybuchowym Ex zgodnych z wymaganiami ATEX dost pne s w j zyku angielskim...

Страница 4: ...cations 8 2 9 Regulatory standards 9 3 INSTALLATION OF 21 FU20 FTS MTS 21 3 1 Typical installation 21 3 2 Preparing the sensor for use 21 3 3 Mounting the sensor 21 3 4 Mounting FU20 FTS MTS using qui...

Страница 5: ...FU20 FTS MTS comes with a multipole VP connector and requires separate cable type WU10 V D or WE10 H D that are available in several lengths Additionally the models type VP and VS can be mounted in al...

Страница 6: ...cess environmental conditions that may be related to the failure of the sensor Statement as to whether warranty or non warranty service is requested Complete shipping and billing instructions for retu...

Страница 7: ...a 2 of theoretical value Note The temperature sensor included in the FU20 FTS MTS is designed for process compensation and for indication It is NOT designed for process temperature control 2 4 Dynamic...

Страница 8: ...L x W x H 300 x 100 x 75 mm 11 8 x 3 9 x 3 0 inch Package weight max 0 33 kg 0 73 lbs 2 7 Environmental conditions Storage temperature 10 to 50 C 14 to 122 F Ingress Protection IP67 conform IEC 60529...

Страница 9: ...e maximum process pressure LVD Directive 2014 35 EU Applying standards NEN EN IEC 61010 1 ANSI ISA 61010 1 CAN CSA C22 2 No 61010 1 Note 1 Models without ID chip with fixed cable or VP type I O signal...

Страница 10: ...ame certificate number is created with a new issue number ATEX DEKRA11ATEX0014 X issue 1 effective from 03 2011 DEKRA11ATEX0014 X issue 2 effective from 02 2021 IECEx IECEx DEK 11 0064X issue 0 effect...

Страница 11: ...rameters Ui 18 V Ii 170 mA Pi 0 4 W Li 0 1 mH models with fixed cable or Li 0 mH VS VP type Ci 150 nF models with fixed cable or Ci 0 4 nF VS type or Ci 0 nF VP type Ambient temperature 40 C to 40 C f...

Страница 12: ...12 IM 12B06J03 05EN P Control drawing D E 2020 023 A50 part 1...

Страница 13: ...Option Codes FU20 ab cd efg h ab Connection type Two alphanumeric characters identifying the length of the permanent cable each character from 0 to 9 VP Connector without ID chip VS Connector with ID...

Страница 14: ...14 IM 12B06J03 05EN P Control drawing D E 2020 023 A50 part 2 SA11 P2...

Страница 15: ...uired isolation 5 Sensor Model code Table 4 Model Suffix Codes Option Codes FU20 ab cd efg h ab Connection type VS Connector with ID chip cd Temperature sensor Region T1 Pt1000 IS for ATEX IECEx FM US...

Страница 16: ...8V Ii 170 mA Pi 0 4 W Li 0 1 mH models with fixed cable or Li 0 mH VS VP type Ci 150 nF models with fixed cable or Ci 0 4 nF VS type or Ci 0 nF VP type Ambient temperature 40 C to 40 C for temperature...

Страница 17: ...17 IM 12B06J03 05EN P Control drawing D E 2020 023 A51 part 1...

Страница 18: ...ID chip cd Temperature sensor Region T1 Pt1000 IS for ATEX IECEx FM US FM CAN efg Type FTS PVDF body Tapered Thread Dome shaped Sodium ions sensitive membrane Silicon Viton sealings MTS PVDF body Tape...

Страница 19: ...19 IM 12B06J03 05EN P Control drawing D E 2020 023 A51 part 2 SA11 P2...

Страница 20: ...d Temperature sensor Region T1 Pt1000 IS for ATEX IECEx FM US FM CAN efg Type FTS PVDF body Tapered Thread Dome shaped Sodium ions sensitive membrane Silicon Viton sealings MTS PVDF body Tapered Threa...

Страница 21: ...tal plane to eliminate air bubbles in the pH glass bulb see Figure 3 Figure 3 Sensor installation 3 2 Preparing the sensor for use Remove the sensor from its shipping box and slide of the so called we...

Страница 22: ...n the process connection Tighten the adapter using a wrench on the adapter flats 3 Apply Teflon tape to the appropriate threaded end of the sensor 4 Place the O ring and screw the mating part of the a...

Страница 23: ...Do not undo the part in the metal tube 5 Lead the sensor cable through the tube of the fitting from the side where the knurled knob has been removed Attach the sensor and cable as usual 6 Hold the se...

Страница 24: ...2B06J03 05EN P Figure 7 Installation examples for the FU20 Fitting O ring Adapter FPS FU20 O ring FU20 FD40 Cleaning solution water acid etc K1547PH K1547PF 60mm HCNF SPRAY CLEANER FU20 adapter Proces...

Страница 25: ...N P 4 DIMENSIONS Dimensions in mm inches Figure 8 Dimensions FU20 FTS MTS Figure 9 Dimensions 1 FU20 FTS MTS adapter Stainless Steel Titanium and FU20 FTS MTS adapter for FF40 FS40 and FD40 fittings N...

Страница 26: ...nnector layout Table 7 Figure 12 Connection diagram connector type Note Preferred connection cable is Yokogawa Model WU10 V D or Model WE10 Pin no Description VP type Description VS type A pH pH B Ref...

Страница 27: ...just the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2 to do a 2 point calibration 4 Rinse sensor thoroughly with tap water 5 Immerse the sensor...

Страница 28: ...ng the complete sensor with solution Some cleaning solutions will damage the model code sticker and connector which are placed on top of the sensor Only clean the measuring elements at the bottom side...

Страница 29: ...t T1 T2 Pt1000 Pt100 n a for FTD and VS Model NPT FSM FTD FTS MTS Dome shape model heavy duty Flat surface model self cleaning Dome shape model chemical resistant Differential pH sensor Differential p...

Страница 30: ...K1547PG Hastelloy nozzle and mounting set HCNF K1547PH Nylon tube 10 meter and tube mounting set for chemical cleaning system K1520BF Buffer solutions Buffer solution pH 4 7 9 pNa 0 500 ml each ionic...

Страница 31: ...L COMPATIBILITY CHART Table 11 Note Information in this list is based on our general experience and literature data and given in good faith However Yokogawa is unable to accept responsibility for clai...

Страница 32: ...32 IM 12B06J03 05EN P IM 12B06J03 05EN P Printed in The Netherlands 01 2101 Subject to change without notice Copyright...

Отзывы: