11
IM 12B06J03-05EN-P
FM-United States
Applying standards : FM Class 3600
FM Class 3610
FM Class 3810
ANSI/ISA 60079-0
ANSI/ISA 60079-11
Certificate no.*
: FM20US0123X
IS CL I, DIV 1, GP ABCD, T3…T6
CL I, ZN 0, AEx ia IIC, T3…T6 Ga
Control Drawing: D&E 2020-023-A50
Electrical data
: See Note 3
Specific conditions : See Control Drawing D&E 2020-023-A50
of use
Note 3:
Intrinsically safe, entity, for Class I, Division 1, Groups A, B, C and D;
Class I, Zone 0, AEx ia IIC, Ga (entity) for hazardous (classified) locations when
installed per control drawing D&E 2020-023-A50.
Sensor input parameters:
Ui= 18 V; Ii= 170 mA; Pi= 0.4 W;
Li= 0.1 mH (models with fixed cable) or Li= 0 mH (VS/VP type);
Ci= 150 nF (models with fixed cable) or
Ci= 0.4 nF (VS type) or Ci= 0 nF (VP type).
Ambient temperature:
-40 °C to +40 °C for temperature class T6,
-40 °C to +55 °C for temperature class T4 and T5,
-40 °C to +105 °C for temperature class T3.
When the sensor has been connected to non intrinsically safe equipment which
exceeds the restrictions regarding the sensor input circuits, the sensor is not
suitable anymore for intrinsically safe use.
* Certification is subject to change, due to new regulations or changes in the product itself.
When a certificate is updated, a new revision under the same certificate number is created
with a new date.
-
FM-United States:
FM20US0123X (effective from 03-2021)
Содержание FU20-FTS
Страница 1: ...IM 12B06J03 05EN P 1st edition Instruction Manual Model FU20 FTS MTS Differential pH ORP sensor...
Страница 12: ...12 IM 12B06J03 05EN P Control drawing D E 2020 023 A50 part 1...
Страница 14: ...14 IM 12B06J03 05EN P Control drawing D E 2020 023 A50 part 2 SA11 P2...
Страница 17: ...17 IM 12B06J03 05EN P Control drawing D E 2020 023 A51 part 1...
Страница 19: ...19 IM 12B06J03 05EN P Control drawing D E 2020 023 A51 part 2 SA11 P2...