IM 12B06J03-05EN-P 

 

 

 

6. GENERAL CALIBRATION & MAINTENANCE PROCEDURE 

 

Calibration of the FU20-FTS/MTS sensor has to be done with the pH analyzer connected. 
Normally the pH standards that are preprogrammed in the pH analyzer can be used for 
calibration with the pH analyzer set to 

“AUTOCAL”

. The FU20-FTS/MTS however is a 

differential pH sensor which needs pH buffers that have the same ionic strength because 
the sodium reference will change as the ionic strength changes. These pH buffers are not 
preprogrammed in the pH analyzer, which means the calibration has to be done with 
specific buffer solutions (see Section 6.1) and the pH analyzer set to 

“MANUAL”

 

calibration. Refer to the analyzer Instruction Manual for details. Calibration data of the 
sensor is stored into the pH analyzer. 
 

6.1 Calibration for pH measurement  

To calibrate the FU20-FTS/MTS sensor, two buffer solutions with known pH values are 
required. It is recommended that one buffer solution has a value near to pH 7.00. 
Depending on the process value to be measured, the second buffer solution should be 
either acidic (below pH 7.00) or alkaline (above pH 7.00). Buffers which are available are: 
pH 4.00, pH 7.00 and pH 9.00. See Section 8 for ordering information. 

 

 
The following is a very general 2-point calibration procedure:  
1. Clean the sensor using a 5% solution of HCl;  
2. Rinse sensor thoroughly with tap water (

DO NOT 

use demineralized water); 

3. Immerse the sensor in the first buffer solution (pH 7.00 is recommended); 
    Set the pH analyzer to 

“MANUAL” 

calibration. Wait until the pH reading is stable.  

    Adjust the pH reading in the pH analyzer to the value indicated on the bottle (in this 
    case 7.00). Go to solution 2 to do a 2-point calibration; 
4. Rinse sensor thoroughly with tap water; 
5. Immerse the sensor in the second buffer (pH 4.00 is recommended); Wait until the pH  
    reading is stable. Adjust the pH reading in the pH analyzer to the value indicated on the  
    bottle (in this case 4.00). Quit the manual calibration routine by accepting the new  
    calculated calibration data of zero and slope.  
6. After calibration, re-install the sensor into the process. 

 

Note: 

It is important to understand it is possible the FU20-FTS/MTS differential sensor 

          does not show the correct pH value in process after buffer calibration. The reason is 
          that the ionic strength of the buffers is 1 mol NaCl, which can deviate from the ionic  
          strength of the process. For an accurate pH reading an extra 1 point process  
          calibration has to be done (see Section 6.2) to compensate for differences in ionic  
          strength. 
 
During calibration, the temperature compensation should be active. The EXA/FLXA 
analyzer automatically compensates for the sensitivity change of the pH sensor at different 
temperatures. After calibration, replace or re-install the sensor into the process.  

 

6.2 Process calibration 

The following is a specific 1-point manual calibration procedure for the process to 
measure: Set the pH analyzer to 

“MANUAL” 

calibration. Wait until the pH reading is 

stable. Adjust the pH reading in the pH analyzer to the value of the process sample that is 
analyzed using a laboratory measurement. For the laboratory measurement, we advise to 
use the Yokogawa PH72 personal pH meter. Quit the manual calibration routine by 
accepting the new calculated calibration data of zero. Slope value will be unchanged. 

Содержание FU20-FTS

Страница 1: ...IM 12B06J03 05EN P 1st edition Instruction Manual Model FU20 FTS MTS Differential pH ORP sensor...

Страница 2: ...oole E Todos los manuales de instrucciones para los productos antiexplosivos de ATEX est n disponibles en ingl s Si desea solicitar las instrucciones de estos art culos antiexplosivos en su idioma loc...

Страница 3: ...com a delega o mais pr xima ou com um representante da Yokogawa PL Wszystkie instrukcje obs ugi dla urz dze w wykonaniu przeciwwybuchowym Ex zgodnych z wymaganiami ATEX dost pne s w j zyku angielskim...

Страница 4: ...cations 8 2 9 Regulatory standards 9 3 INSTALLATION OF 21 FU20 FTS MTS 21 3 1 Typical installation 21 3 2 Preparing the sensor for use 21 3 3 Mounting the sensor 21 3 4 Mounting FU20 FTS MTS using qui...

Страница 5: ...FU20 FTS MTS comes with a multipole VP connector and requires separate cable type WU10 V D or WE10 H D that are available in several lengths Additionally the models type VP and VS can be mounted in al...

Страница 6: ...cess environmental conditions that may be related to the failure of the sensor Statement as to whether warranty or non warranty service is requested Complete shipping and billing instructions for retu...

Страница 7: ...a 2 of theoretical value Note The temperature sensor included in the FU20 FTS MTS is designed for process compensation and for indication It is NOT designed for process temperature control 2 4 Dynamic...

Страница 8: ...L x W x H 300 x 100 x 75 mm 11 8 x 3 9 x 3 0 inch Package weight max 0 33 kg 0 73 lbs 2 7 Environmental conditions Storage temperature 10 to 50 C 14 to 122 F Ingress Protection IP67 conform IEC 60529...

Страница 9: ...e maximum process pressure LVD Directive 2014 35 EU Applying standards NEN EN IEC 61010 1 ANSI ISA 61010 1 CAN CSA C22 2 No 61010 1 Note 1 Models without ID chip with fixed cable or VP type I O signal...

Страница 10: ...ame certificate number is created with a new issue number ATEX DEKRA11ATEX0014 X issue 1 effective from 03 2011 DEKRA11ATEX0014 X issue 2 effective from 02 2021 IECEx IECEx DEK 11 0064X issue 0 effect...

Страница 11: ...rameters Ui 18 V Ii 170 mA Pi 0 4 W Li 0 1 mH models with fixed cable or Li 0 mH VS VP type Ci 150 nF models with fixed cable or Ci 0 4 nF VS type or Ci 0 nF VP type Ambient temperature 40 C to 40 C f...

Страница 12: ...12 IM 12B06J03 05EN P Control drawing D E 2020 023 A50 part 1...

Страница 13: ...Option Codes FU20 ab cd efg h ab Connection type Two alphanumeric characters identifying the length of the permanent cable each character from 0 to 9 VP Connector without ID chip VS Connector with ID...

Страница 14: ...14 IM 12B06J03 05EN P Control drawing D E 2020 023 A50 part 2 SA11 P2...

Страница 15: ...uired isolation 5 Sensor Model code Table 4 Model Suffix Codes Option Codes FU20 ab cd efg h ab Connection type VS Connector with ID chip cd Temperature sensor Region T1 Pt1000 IS for ATEX IECEx FM US...

Страница 16: ...8V Ii 170 mA Pi 0 4 W Li 0 1 mH models with fixed cable or Li 0 mH VS VP type Ci 150 nF models with fixed cable or Ci 0 4 nF VS type or Ci 0 nF VP type Ambient temperature 40 C to 40 C for temperature...

Страница 17: ...17 IM 12B06J03 05EN P Control drawing D E 2020 023 A51 part 1...

Страница 18: ...ID chip cd Temperature sensor Region T1 Pt1000 IS for ATEX IECEx FM US FM CAN efg Type FTS PVDF body Tapered Thread Dome shaped Sodium ions sensitive membrane Silicon Viton sealings MTS PVDF body Tape...

Страница 19: ...19 IM 12B06J03 05EN P Control drawing D E 2020 023 A51 part 2 SA11 P2...

Страница 20: ...d Temperature sensor Region T1 Pt1000 IS for ATEX IECEx FM US FM CAN efg Type FTS PVDF body Tapered Thread Dome shaped Sodium ions sensitive membrane Silicon Viton sealings MTS PVDF body Tapered Threa...

Страница 21: ...tal plane to eliminate air bubbles in the pH glass bulb see Figure 3 Figure 3 Sensor installation 3 2 Preparing the sensor for use Remove the sensor from its shipping box and slide of the so called we...

Страница 22: ...n the process connection Tighten the adapter using a wrench on the adapter flats 3 Apply Teflon tape to the appropriate threaded end of the sensor 4 Place the O ring and screw the mating part of the a...

Страница 23: ...Do not undo the part in the metal tube 5 Lead the sensor cable through the tube of the fitting from the side where the knurled knob has been removed Attach the sensor and cable as usual 6 Hold the se...

Страница 24: ...2B06J03 05EN P Figure 7 Installation examples for the FU20 Fitting O ring Adapter FPS FU20 O ring FU20 FD40 Cleaning solution water acid etc K1547PH K1547PF 60mm HCNF SPRAY CLEANER FU20 adapter Proces...

Страница 25: ...N P 4 DIMENSIONS Dimensions in mm inches Figure 8 Dimensions FU20 FTS MTS Figure 9 Dimensions 1 FU20 FTS MTS adapter Stainless Steel Titanium and FU20 FTS MTS adapter for FF40 FS40 and FD40 fittings N...

Страница 26: ...nnector layout Table 7 Figure 12 Connection diagram connector type Note Preferred connection cable is Yokogawa Model WU10 V D or Model WE10 Pin no Description VP type Description VS type A pH pH B Ref...

Страница 27: ...just the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2 to do a 2 point calibration 4 Rinse sensor thoroughly with tap water 5 Immerse the sensor...

Страница 28: ...ng the complete sensor with solution Some cleaning solutions will damage the model code sticker and connector which are placed on top of the sensor Only clean the measuring elements at the bottom side...

Страница 29: ...t T1 T2 Pt1000 Pt100 n a for FTD and VS Model NPT FSM FTD FTS MTS Dome shape model heavy duty Flat surface model self cleaning Dome shape model chemical resistant Differential pH sensor Differential p...

Страница 30: ...K1547PG Hastelloy nozzle and mounting set HCNF K1547PH Nylon tube 10 meter and tube mounting set for chemical cleaning system K1520BF Buffer solutions Buffer solution pH 4 7 9 pNa 0 500 ml each ionic...

Страница 31: ...L COMPATIBILITY CHART Table 11 Note Information in this list is based on our general experience and literature data and given in good faith However Yokogawa is unable to accept responsibility for clai...

Страница 32: ...32 IM 12B06J03 05EN P IM 12B06J03 05EN P Printed in The Netherlands 01 2101 Subject to change without notice Copyright...

Отзывы: