<2. WIRING AND INSTALLATION>
2-24
IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00
2.6.6
Wiring sensors with SENCOM SA
When suffix code -S5 (SENCOM SA) for 1st or 2nd input is specified.
The next figure shows a wiring with WU11, which connects to SENCOM SA.
Be sure to find the label “SENCOM SA” beside the terminals.
Screws are all M3 with washers.
Figure 2.29
SENCOM module terminal assignment
82
86
83
84
87
SHLD
GND
B-
A+
V5
Signal ground
Data (negative)
Data (positive)
Power supply
for SENCOM SA
WU11 cable
Multiple
connection
SA11
Sensor
Sensor
Sensor
SA11
SA11
BA11
WE10
Extension cable
*1
*2
WU11
cable
Figure 2.30
Wiring SENCOM SA
*1:
Multi sensor connection: 1st sensor module can connect up to 4 sensors equipped with SA11 SENCOM SA, via BA11 junction
box.
*2:
WE10 is an extension cable stretching between SA11 and sensors. WE10 is universal to connections if needed.
Connect a SA11 SENCOM Smart adapter to a sensor. Connect SA11 to FLXA 402 via WU11
SENCOM cable.
Read each exclusive user’s manual for information on SA11 Smart Adapter, WU11
Interconnection/Extension Cable for SENCOM Products, WE10 extension cable, BA11 Active
Junction Box.
Because WU11 cables are thin, first remove the cable gland’s standard packing and apply the
supplied grommet instead.
Remove the standard packing.
Sensor
Grommet
FLXA402 acknowledges the connection of sensors, if not, FLXA402 alerts an error.