<2.  WIRING AND INSTALLATION>

2-14

IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00

n

  mA input

This describes a wiring when the suffix code for mA Output/Input is -A4.
FLXA 402 receives 4-20 mA analog signals inputs of pressure compensation for DO sensors, or 

of temperature compensation for pH, SC, ISC, DO sensors.
Use shielded cable for input signal cable and connect the shield to terminal 89. Rated voltage 

of cable is 30 V AC or greater. For wiring use multicore shielded cable. Screws for IO module 

terminals are all M3 with washers.

Terminal 

number

Contact 

terminals

Description

87

AI+

4-20 mA input +

88

AI--

4-20 mA input-

89

SHILD

Shield

Figure 2.14 

IO module terminal assignment

n

  Contact input

IO module contains contact input function which is for starting WASH CYCLE or changing 

RANGE of 4-20 mA output.
For input signal cable, use shielded cables. Connect the shield to terminal 63.
All screws are M3 with washers.

Terminal 

number

Contact 

terminals

Description

21

Dl1

Contact input 1

22

COM

Contact input common

24

DI2

Contacct input 2*1

63

SHILD

Shield

*1 

when mA IO code -A4 is selected

Figure 2.15 

IO module terminal assignment

When all wiring is complete, close the FLXA402 front cover, and turn the power on. 
Check that the screen displays normally.

CAUTION

To install the front panel, tighten each screw evenly until every four of them is completely 

fastened. For example turn each screw in two times and repeat it until all of the screws are evenly 

tightened up
Do not use an electric screwdriver with high revolutions such as 1000 rpm. The revolutions of the 

electric screwdriver should be less than 400 rpm.
Four front panel screws should be tightened to the following torque; 1.5 to 1.6 N•m

Содержание FLXA402

Страница 1: ...User s Manual Model FLXA402 4 Wire Converter Installation and Wiring Introduction and General Description IM 12A01F01 02EN IM 12A01F01 02EN 2nd Edition...

Страница 2: ...N Online manual FLXA402 4 Wire Converter Operation of pH ORP IM 12A01F02 01EN Online manual FLXA402 4 Wire Converter Operation of SC IM 12A01F03 01EN Online manual FLXA402 4 Wire Converter Operation o...

Страница 3: ...Composition of this User s Manual FLXA402 a modular designed converter is a multi parameter instrument offering a wide range of measurement choices such as pH ORP oxidation reduction potential Resist...

Страница 4: ...merique de la classe A est conforme a la norme NMB 003 du Canada This device complies with Industry Canada s applicable license exempt RSSs Operation is subject to the following two conditions 1 this...

Страница 5: ...Yokogawa Electric Corporation hereby declare that this equipment FLXA402 is in compliance with the essential requirements and other relevant provisions of Directive 2014 53 EU The full text of the EU...

Страница 6: ...s 1 10 2 WIRING AND INSTALLATION 2 1 2 1 Installation site 2 1 2 2 Mounting methods 2 2 2 3 Removing the shield cover 2 3 2 4 Installing the cable glands 2 4 2 5 Wiring 2 6 2 5 1 Wiring the power supp...

Страница 7: ...Blank Page...

Страница 8: ...upports the HART and MODBUS protocols This solution also facilitates connection to the cloud and the construction of an Industrial IoT IIoT environment for improved field maintenance For further infor...

Страница 9: ...d then immediately notify the carrier and the relevant Yokogawa sales office n Checking the model and suffix code l FLXA402 Make sure the model and suffix code on the nameplate affixed to the left sid...

Страница 10: ...ne below Cable gland M20x1 5 8 pcs Rubber plug attachment 5 pcs digital communication E Modbus TCP IP Cable gland M20x1 5 7 pcs Ethernet cable gland black 1 piece Rubber plug attachment black 5 pcs Gr...

Страница 11: ...ption code UM U 1 set For UM bracket for U and PM one of each are supplied Panel mounting hardware Option code UM PM 1 set Hood Option code H6 H7 1 set Stainless steel tag plate Option code SCT 1 set...

Страница 12: ...4 pcs Cable gland for adapter 4 pcs Rubber plug attachment 1 piece CB6 CD6 CF6 Adapter 3 pcs Cable gland for adapter 3 pcs Ethernet adapter 1 piece Ethernet cable gland black 1 piece Rubber plug atta...

Страница 13: ...Front view with the front panel opened Reboot switch Do not use it Terminal block for power supply LCD display Slide switch for HART only Front panel Module Figure 1 2 FLXA402 Parts names and descrip...

Страница 14: ...fied for 1st 2nd sensor module defines what is printed on the shield cover next to the sensor module Figure 1 3 shows an example of selecting P1 for both modules The following list shows what is print...

Страница 15: ...jumpers store them in the jumper holders located inside of the front panel to prevent them from being lost You can put jumpers anywhere among 5 holders 1A and 4Bs as Figure 1 4 A has a guide hole B h...

Страница 16: ...NN Option Mounting hardware Hood Tag plate Conduit adapter 7 UM U PM H6 H7 SCT CB4 CD4 CF4 CB6 CD6 CF6 Universal mounting kit 5 Pipe and wall mounting hardware Panel mounting hardware Hood stainless s...

Страница 17: ...all mounting stainless K9703SS 1 set same as Option code U for panel mounting stainless K9703ZD 1 set same as Option code PM Sun shade hood Stainless K9698WK 1 set same as Option code H6 stainless ure...

Страница 18: ...ambient temperature and humidity are within the limits of the instrument specifications as below If the instrument is installed outdoors and exposed to direct sunlight a sun shade hood should be used...

Страница 19: ...is used for the wall mounting Note The wall on which the analyzer is mounted should be strong enough to bear the weight of more than 8kg Note Horizontal pipe mounting is possible by turning the bracke...

Страница 20: ...shut down the power supply before you remove the shield cover Open the front panel and remove the shield cover After removing the shield cover install the cable glands To remove the shield cover loos...

Страница 21: ...the modules for communication For installing conduit adapters apply Conduit adapter set need to be purchased additionally seel Conduit adapter need to be purchased additionally l Conduit adapter need...

Страница 22: ...es which conduit adapters connect to instead of using supplied accessory cable gland dark gray Figure 2 5 Caps of cable glands are not used For Ethernet wiring use Cable gland for Ethernet black and C...

Страница 23: ...d cover After the wiring screw up the shield cover WARNING Use wiring cables with heat resistance above 75 C waterproof rating IP66 or higher Four front panel screws should be tightened to the followi...

Страница 24: ...1 or IEC60947 3 Yokogawa recommend installing the external power supply switch circuit breaker and FLXA402 converter all in the same location Install the external power supply switch or circuit breake...

Страница 25: ...de with those of the cables Use Conductors with a cross sectional area of 1 25 mm2 or greater Cables with an outer diameter of 6 to 12 mm Cables with rated voltage 30 V AC or greater Figure 2 7 DC pow...

Страница 26: ...ternal grounding NOTE FLXA402 on AC power supply requires the protective grounding The converter on DC power supply does not require the protective grounding but it does function grounding It has the...

Страница 27: ...A402 has 4 contact outputs which allows configuration to adjust to operation Contact S4 is fail safe contact See IM Converter operation IM 12A01F01 03EN Section 4 4 There are 2 cable inlets Take a sui...

Страница 28: ...92 B data negative 91 A data positive 95 Term terminal for termination resistor 110 Figure 2 10 Network module terminal Use multi core shielded cable whose core wire is twisted pair Connect them to te...

Страница 29: ...fied Modbus TCP IP Figure 2 12 shows the positions of RJ45 Use STP shield cable of category 5 or greater to the RJ45 connector You can apply both straight and crossover cable for Ethernet cabling Modb...

Страница 30: ...s for output signal cables and connect shield to terminal 63 or 89 When mA I O is A4 the terminal 89 is used for mA input and cannot be used for output Use cables of rating 30 V AC or greater For wiri...

Страница 31: ...tarting WASH CYCLE or changing RANGE of 4 20 mA output For input signal cable use shielded cables Connect the shield to terminal 63 All screws are M3 with washers Terminal number Contact terminals Des...

Страница 32: ...in diameter max 1 9 mm Ring and fork spade terminal width max 7 8 mm For details on the sensors refer to the respective instruction manuals of the sensors NOTE When the second input is used the upper...

Страница 33: ...14 15 16 Fixed Cable PR20 PR10 pH 11 12 13 14 15 16 Yes No Fixed Cable SC21 pH Blue Liquid Earth Red White No No WU20 Triax SC24V pH E F C H A B D VP8 DC SC25V pH E F B D A Yes No VP6 SC SC29 PTG pH C...

Страница 34: ...SC INCHTRODE pH E F B D A Yes No VP6 SC MECOTRODE pH Blue Red White Yes No 13 14 WU20D Cable MECOTRODE VP pH E F B A Yes No VP8 DC OXYFERM VP OXYGOLD DO White Green Green Yellow Black Red Hamilton VP...

Страница 35: ...jumpers are supplied with each module Keep unused jumpers in the jumper holder to avoid losing them pH Glass Electrode is a high impedance sensor input Standard reference electrodes and an ORP REDOX...

Страница 36: ...ollows 11 Temperature 12 Temperature 13 Reference 14 Solution ground 15 Metal measure 16 Shield 17 Shield FLXA402 REF TC Liquid Earth ORP For conventional ORP sensors the jumpers are placed as illustr...

Страница 37: ...will accommodate the cable adequately Single cables between approximately 6 mm and 12 mm 0 24 and 0 47 can be sealed properly with these glands and the standard tule Remove standard packing Sensor Gr...

Страница 38: ...6 3 Wiring the inductive conductivity ISC sensor ISC40 sensors are connected to the module as follows 11 Temp Receive coil Drive coil 12 13 17 15 16 14 shield Sensor shield internal shield FLXA402 The...

Страница 39: ...ent types of DO sensors i Galvanic sensors like model DO30G ii Polarographic sensors like HAMILTON S Oxyferm and Oxygold The connection is as follows 11 TC TC anode galvanic cathode galvanic anode pol...

Страница 40: ...any terminal of somewhere between 15 to 18 14 13 12 11 10 SHLD Exclusive cable Green and Yellow Brown Yellow White Green Red Transparent Black Transparent FLXA402 terminal number GND B A 12V DO70G Fig...

Страница 41: ...U11 cable Figure 2 30 Wiring SENCOM SA 1 Multi sensor connection 1st sensor module can connect up to 4 sensors equipped with SA11 SENCOM SA via BA11 junction box 2 WE10 is an extension cable stretchin...

Страница 42: ...Revision Record Manual Title Model FLXA402 4 Wire Converter Installation and Wiring Introduction and General Description Manual No IM 12A01F01 02EN Feb 2019 2nd Edition Modified Model Suffix Codes p...

Страница 43: ...Blank Page Yokogawa Electric Corporation 2 9 32 Nakacho Musashino shi Tokyo 180 8750 JAPAN http www yokogawa com...

Отзывы: