<2. WIRING AND INSTALLATION>
2-5
IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00
l
Conduit adapter
When protecting the cable with a conduit, use an adapter. (option codes: /CB4, /CD4, /CF4, /
CB6/ CD6 /, /CF6)
CAUTION
There are 8 holes for cable connection. Apply the cable gland of supplied accessories (dark-
grayed cable glands) to the holes which no conduit adapters connect to.
Apply the supplied cable gland for adapter (white) to the holes which conduit adapters connect to
instead of using supplied-accessory cable gland (dark gray). (Figure 2.5)
Caps of cable glands are not used.
For Ethernet wiring, use Cable gland for Ethernet (black) and Conduit adapter for Ethernet
(punched with a letter “E”).
G1/2 (/CB4, /CB6)
1/2NPT (/CD4, /CD6)
M20x1.5 (/CF4, /CF6)
Adapter
Cable gland
for adapter
Case
Case
G1/2 (/CB6)
1/2NPT (/CD6)
M20x1.5 (/CF6)
Adapter
for Ethernet
Cable gland
for Ethenet
Figure 2.5
Conduit adapter (option)
CAUTION
When using a cable conduit, use a flexible conduit to avoid stress on the conduit adapter.
The excessive stress on the conduit adapter may damage the housing.
Содержание FLXA402