<2. WIRING AND INSTALLATION>
2-15
IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00
2.6
Sensor wiring
The FLXA402 can be used with a wide range of commercially available sensor types, both from
Yokogawa and other manufacturers.
Terminal screw size is M3, and torque of screw up is 0.6 N•m.
Pin terminal, ring terminal and spade terminal can be used.
Pin terminal:
pin diameter: max. 1.9 mm
Ring and fork (spade) terminal: width:
max. 7.8 mm
For details on the sensors, refer to the respective instruction manuals of the sensors.
NOTE
When the second input is used, the upper-level module is for the first input. Be careful not to
make a mistake when wiring the modules.
If the second input is available, wire the second module in the lower level first.
There are two cable entry holes. Either one can be used.
When the second input s not available, be sure to plug holes of unused cable gland with the
supplied rubber close up plug.
Содержание FLXA402