75
IM 04L45B02-02EN
Installation and W
iring
Recommended Replacement
Periods for W
orn Parts
Setup Items and
Default V
alues
Connecting to the RS-422/485 (/C3)
FG (Frame Ground)
Frame ground of the DX.
SG (Signal Ground)
Signal ground.
SDB (Send Data B)
Send data B (+).
SDA (Send Data A)
Send data A (–).
RDB (Received Data B)
Receive data B (+).
RDA (Received Data A)
Receive data A (–).
FG SG SDB SDA RDB RDA
Electric
potential
of the shield
Two-wire system
Shield
FG SG SDB SDA RDB RDA
Electric
potential
of the shield
Four-wire system
Shield
Recommended length of stripped wire: 9 mm. Recommended tightening torque: 0.4-0.5 N•m.
Cable
There are two types of cables available, the four-wire cable and the two-wire cable (used only for the Modbus protocol).
The cable should meet the following specifications.
• Type:
Shielded twisted pair cable: 3 pairs 24 AWG or more (four-wire),
2 pair 24 AWG or more (two-wire)
• Characteristic impedance: 100 Ω
• Capacitance:
50 pF/m
• Total cable length:
Up to 1.2 km
Connecting to the VGA Connector (/D5)
CAUTION
• Connect the cable after turning OFF the DX and the monitor.
• Do not short the VIDEO OUT terminal or apply external voltage to it. If you do,
the DX may malfunction.
VGA Output Terminal
!
VIDEO OUT(VGA)
The DX display can be output to a monitor through the RGB output. Connectable
monitors are VGA monitors or multi-sync monitors capable of displaying VGA.
Specifications of the VGA Output Terminal
Pin No. Signal Name
Specifications
1
Red
0.7 Vp-p
2
Green
0.7 Vp-p
3
Blue
0.7 Vp-p
4
–
5
–
6
GND
7
GND
8
GND
9
–
10
GND
11
–
12
–
13
Horizontal sync
signal
Approx. 31.5 kHz, TTL
negative logic
14
Vertical sync signal
Approx. 60 Hz, TTL
negative logic
15
–